COMMONWEALTH vs. KIRK P. CAMBLIN.
SJC-11774
Supreme Judicial Court of Massachusetts
September 7, 2017. - December 8, 2017.
Gants, C.J., Lenk, Gaziano, Budd, Cypher, & Kafker, JJ.
Middlesex. NOTICE: All slip opinions and orders are subject to formal revision and are superseded by the advance sheets and bound volumes of the Official Reports. If you find a typographical error or other formal error, please notify the Reporter of Decisions, Supreme Judicial Court, John Adams Courthouse, 1 Pemberton Square, Suite 2500, Boston, MA, 02108-1750; (617) 557-1030; SJCReporter@sjc.state.ma.us
Motor Vehicle, Operating under the influence. Evidence, Breathalyzer test, Scientific test.
Complaint received and sworn to in the Ayer Division of the District Court Department on April 28, 2008.
Following review by this court, 471 Mass. 639 (2015), a motion to exclude evidence as scientifically unreliable was heard by Mark A. Sullivan, J.
Andrew W. Piltser Cowan for the defendant.
Casey E. Silvia, Assistant District Attorney (Cyrus Y. Chung & Laura S. Miller, Assistant District Attorneys, also present) for the Commonwealth.
GAZIANO, J. In Commonwealth v. Camblin, 471 Mass. 639, 640, 651 (2015) (Camblin I), we remanded this case to the District Court to conduct a hearing on the scientific reliability of a particular model of breathalyzer, the Alcotest 7110 MK III-C (Alcotest), while retaining jurisdiction of the case. After conducting a Daubert-Lanigan hearing, a District Court judge found that the Alcotest was capable of producing scientifically reliable
1. Background. a. Prior proceedings.
In 2008, a District Court complaint issued charging the defendant with operating a motor vehicle while under the influence of liquor (OUI), in violation of
The defendant then filed a petition pursuant to
In Camblin I, supra, the defendant primarily challenged the reliability of the Alcotest on the ground of asserted errors in the source code for its computer programs. In doing so, the defendant relied upon, among other things, affidavits from two expert witnesses and a report that he had submitted in support of his motion to exclude. One of the experts averred that he had scanned the Alcotest‘s source code, utilizing an “industry standard source code analysis tool,” and had found more than 7,000 errors and 3,000 warning signals. Id. at 651. A different expert averred that the Alcotest is incapable of measuring accurately the amount of ethanol in a breath sample because the device does not exclude other “interfering substances” that might be present in the sample.2 Id. at 652-653.
In addition, the report that the defendant submitted suggested that the calibration test used with the Alcotest does not ensure accurate results. Id. at 654. We remanded the matter to the District Court for a hearing to consider three issues: (1) the reliability of the Alcotest source code; (2) whether the Alcotest is capable of testing exclusively for ethanol; and (3) whether any source code errors affect the ability of the Alcotest to calculate a subject‘s blood alcohol content (BAC). Id. at 651-655.
b. Proceedings on remand.
On remand, the judge conducted a Daubert-Lanigan hearing, at which experts for the defendant and the Commonwealth testified about the reliability of the Alcotest. After the hearing, the judge issued a decision containing his comprehensive findings of fact, and remitted them to this court. With respect to the defendant‘s challenge to reliability of the source code, the judge concluded that “despite the minor flaws in
In this appeal, the defendant has chosen not to pursue his arguments concerning the source code as the primary basis for the asserted lack of reliability in the Alcotest. Rather, the current focus of the defendant‘s challenge to the reliability of the Alcotest is that it cannot distinguish ethanol from other “interfering” substances that might be present in a breath sample. The following facts were adduced at the Daubert-Lanigan hearing.
The Alcotest is an evidential breath-testing device manufactured by Draeger Safety Diagnostics, Inc. (Draeger). Draeger describes the Alcotest as a “dual sensoric instrument” because it utilizes both infrared spectroscopy and electrochemical fuel cell sampling to analyze alcohol content in a breath sample. The subject blows air into the device through a tube connected to a chamber. An infrared light source at one end of the chamber generates energy in the 9.5 micron range,3 and a detector on the opposite end of the chamber receives the energy from the infrared source. Because infrared energy is absorbed by ethanol molecules, any such molecules that are present in a breath sample effectively “soak up” the infrared energy, and that portion of it does not reach the detector. The Alcotest is designed to measure a subject‘s breath alcohol content based on the amount of infrared energy that reaches the detector as compared to the amount of energy detected when the chamber has been cleared and is filled simply with ambient air. In other words, the reduction in infrared energy (which has been absorbed by the ethanol molecules) from one end of the chamber to the other is equivalent to the concentration of alcohol present in the chamber.
In a dual-sensor Alcotest device, at the same time that the infrared energy is passing through the main chamber, a small portion of the breath sample enters a fuel cell sensor for a second measurement of breath alcohol. The fuel cell is an electrochemical device that essentially operates like a battery. It generates an electrical current from energy produced by a chemical reaction between any ethanol and the oxygen contained within the breath
Both sides presented expert witness testimony on the question whether the Alcotest is capable of testing exclusively for ethanol, while excluding interfering substances. The defendant introduced testimony by Dr. Donald J. Barry, Ph.D., an astronomer with a substantial background in infrared spectrometry technology, as well as a background in chemistry. Barry testified that, where interfering substances are present, the Alcotest is incapable of testing exclusively for ethanol, and therefore its results can be tainted by the presence of interfering substances in the sample. Barry explained that the Alcotest‘s infrared spectroscope identifies a carbon-oxygen molecule that is emitted at a 9.5 micron wavelength. Several organic compounds other than ethanol, including acetone4 and methanol, also emit energy within the 9.5 micron range of the electromagnetic spectrum, and would similarly be detected by the Alcotest‘s spectroscope. Barry concluded that, as a result, the Alcotest could not reliably isolate and identify ethanol in a subject‘s breath to the exclusion of other interfering substances sharing a similar molecular structure.
Barry was not familiar with the particular fuel cell technology used in the Alcotest. He opined generally, however, that, for this type of application, fuel cells are a suspect measuring technology for several reasons, including diminishing performance over time. He testified that he was aware of no scientific support for Draeger‘s assertion that the Alcotest fuel cell sensor is independently capable of detecting alcohol as opposed to other compounds.
The Commonwealth introduced expert testimony by Hansueli Ryser, a Draeger vice-president with a thirty-four year background in engineering evidential breath test devices. Ryser had been involved in the engineering and development of the Alcotest. He testified that there were two primary means by which the
Ryser agreed that some substances, like methanol, do absorb energy at a similar rate and wavelength to ethanol. Nonetheless, Ryser stated that differences in energy absorption rates would allow the Alcotest to distinguish between the two substances, due to the substances’ differences in intensity.
Second, Ryser testified that the fuel cell sensor in an Alcotest machine functions as a fail-safe to distinguish between ethanol and other interfering substances. He noted that the fuel cell measures the flow of electrical current produced by the chemical reaction and registers a “kinetic reactivity” “footprint” for the breath sample. The Alcotest compares this footprint to the footprint created by the flow of electrical current generated by the known ethanol sample contained in a calibrated simulator solution. The Alcotest identifies interfering substances based on whether there are disparities between the two footprints. Finally, the Alcotest compares the infrared spectrometry and electrochemical fuel cell test results. If the two components produce substantially different measurements of a subject‘s BAC level, the Alcotest is designed to flag the differences as caused by an interfering substance, and thereafter to abort the test.
After the hearing,6 the judge concluded that the Alcotest “reliably
2. Discussion. a. Standard of review.
The admission of scientific testimony is governed by what has come to be known as the Daubert-Lanigan standard. Commonwealth v. Senior, 433 Mass. 453, 458 (2001), citing Daubert, 509 U.S. at 585-595, and Lanigan, 419 Mass. at 25-26. See
In Lanigan, 419 Mass. at 25-26, we adopted, in part, the United States Supreme Court‘s reasoning in Daubert, governing the admissibility of expert testimony based on a scientific theory. We did not, however, entirely abandon our prior test to determine the admissibility of scientific evidence, which focused on “whether the community of scientists involved generally accepts the theory or process.” Lanigan, supra at 24, quoting Frye v. United States, 293 F. 1013 (D.C. Cir. 1923). Rather, we held that “general acceptance in the relevant scientific community will continue to be the significant, and often the only, issue,” but “that a proponent of scientific opinion evidence may demonstrate the reliability or validity of the underlying scientific theory or process by some other means.” See Lanigan, supra at 26.
Under the Daubert-Lanigan standard, a judge considering a motion to introduce expert testimony initially considers a nonexclusive list of five factors. See Commonwealth v. Powell, 450 Mass. 229, 238 (2007). Among these factors are “whether the scientific theory or process (1) has been generally accepted in the relevant scientific community; (2) has been, or can be, subjected
b. Reliability of the Alcotest.
Turning to the judge‘s decision in this case, we consider whether he abused his discretion in finding that the Alcotest breathalyzer had been subject to sufficient independent testing to establish its reliability. In reaching his decision that the reliability of the Alcotest had been sufficiently established, the judge relied upon testimony by Ryser and testing conducted by two agencies: the National Highway Traffic Safety Administration (NHTSA), and the Organisation Internationale de Métrologie Légale (OIML), an agency that regulates the use of alcohol breath-testing devices in Europe. The judge also noted that the Alcotest had been approved by the Australian International Laboratory of Spectroscopy and the Forensic Science Academy in Ottawa, Canada.
The judge‘s reliance on NHTSA testing did not constitute an abuse of discretion. “NHTSA certification is widely accepted by courts as evidence of a device‘s reliability.” United States v. Ahlstrom, 530 Fed. Appx. 232, 239 (4th Cir. 2013), citing California v. Trombetta, 467 U.S. 479, 489 & n.9 (1984), and United States v. Brannon, 146 F.3d 1194, 1196 (9th Cir. 1998). See People v. Vangelder, 58 Cal. 4th 1, 33-34 (2013), cert. denied, 134 S. Ct. 2839 (2014) (noting that devices which meet NHTSA evidential breath-testing specifications produce sufficiently reliable results within California‘s regulatory scheme).
Indeed, under
In reaching his determination that “the Alcotest [infrared] and [electrochemical] features were tested independently for their ability to detect non-ethanol substances and both components met the NHTSA specifications,” the judge relied on testimony by Ryser, which he credited explicitly, concerning the two-test comparison standard that the Alcotest employs. The judge noted also that each of the Alcotest‘s two sensors have been found compliant with the NHTSA specifications when operated individually. The judge noted that this independent, dual testing capability further “assures the device‘s ethanol specificity testing capability.” If the results from either of the tests differ by more than a specified threshold, the test is rejected and testing ceases. At the time that the evidentiary hearing was held on remand, no other breathalyzer used a dual-sensor system.
Likewise, the judge‘s reliance on the OIML test specifications and certifications was within his discretion. The OIML‘s certification requirements generally are viewed as being much more stringent than those applicable in the United States. Ryser explained that, because of the wavelength at which it operates, and the precision of its sensor, the Alcotest was even able to meet the certification requirements of the OIML “draft three” set of specifications. The draft three certifications were so stringent that the testing agency itself decided to remove some of those requirements from the “draft four” specification level, the most recent set of certification standards. In large part, the reduced stringency
In sum, the judge was warranted in crediting Ryser‘s testimony that the NHTSA and the OIML certifications further demonstrated that the Alcotest was capable of testing exclusively for ethanol.
The defendant contends that the specification testing conducted by the NHTSA and the OIML was deficient because neither agency utilized mixed samples in evaluating the Alcotest‘s ability to distinguish between ethanol and interfering substances. According to the defendant‘s expert witness, adequate testing requires challenging the device with a wide variety of “physiologically important” substances, as well as studying how the properties of the fuel cell change over time. We discern no error. The judge was well within his discretion in relying upon the NHTSA and OIML reports, based on their standard and widely accepted protocols regarding mixed sample testing. While the judge generally credited Barry‘s testimony, there was no abuse of discretion in the judge‘s decision to reject Barry‘s specific criticisms of the testing methodology. See Canavan‘s Case, 432 Mass. at 312 (judge is qualified to determine questions of credibility concerning proposed scientific expert testimony).
We turn to whether the technology underlying the Alcotest has been subjected to peer review and publication, another of the factors in the Daubert-Lanigan analysis. The peer-review prong of the Daubert-Lanigan standard serves a function similar to the general acceptance test; in essence, it requires a judge to determine whether the scientific theory underlying the disputed evidence has been accepted by the relevant scientific community. See Commonwealth v. Senior, 433 Mass. 453, 460-461 (2001). At the hearing in this case, the Commonwealth submitted a single peer-reviewed article that surveyed the history of breath-testing devices; the article mentioned that the Alcotest‘s use of an infrared detector operating at 9.5 microns and an electrochemical fuel cell “is a highly desirable feature for medicolegal purposes.”8 The judge found that support for the peer-review factor of the Daubert-Lanigan analysis “is admittedly thin.”
In addition to considering whether the Alcotest had been subject to adequate testing and peer review, the judge also found “abundant evidence that the Alcotest and its underlying technology” had satisfied the other nonexclusive factors in the Daubert-Lanigan analysis: it has been generally accepted in the scientific community, it does not have an unacceptably high known or potential rate of error, and it is governed by recognized standards. There was no abuse of discretion in the judge‘s determination that these factors had been met.
3. Conclusion.
The order denying the motion to exclude evidence as scientifically unreliable is affirmed. The judgment of conviction is also affirmed.
So ordered.
Notes
According to the National Safety Council, an “interfering substance” is a “non-ethanol substance” able “to produce a significant response on any breath alcohol testing instrument.” To qualify as an interfering substance, the substance must:
“1. Be a volatile organic compound capable of appearing in the breath of a living, conscious human being.
“2. Be present in sufficiently high concentration to be measured by the instrument after a 15 to 20 minute pretest observation period.
“3. Be able to produce a response on the instrument that is indistinguishable from ethanol.”
National Safety Council, Committee on Alcohol and Other Drugs, Report on the Specificity of Breath Alcohol Analyzers (Feb. 22, 2010).
Draeger offers an optional sensor that measures a subject‘s breath temperature as a means to assist in evaluating breathalyzer test results. Massachusetts law enforcement agencies declined to purchase this option. The defendant contends that the absence of a temperature sensor diminishes the accuracy of the Alcotest, and further demonstrates that it is not reliable. The defendant did not raise this issue in his original appeal from the denial of his motion to exclude the breath test results, and we did not ask the judge to consider it on remand. The judge made no factual finding regarding the significance of installing a breath temperature sensor.
Because the defendant did not raise the issue in his original appeal, we consider it to be waived. See Commonwealth v. Pisa, 384 Mass. 362, 365-367 (1981). We also are persuaded by the New Jersey Supreme Court‘s holding in State v. Chun, 194 N.J. 54, 105-106 (2008), cert. denied, 555 U.S. 825 (2008), that there is insufficient scientific study on the impact of a breath temperature measurement on the accuracy of the results. The New Jersey Supreme Court concluded in that case that the absence of a temperature sensor did not render the Alcotest unreliable. Id. at 107-108. The court commented that, to the extent that variation in breath temperature has an effect on breathalyzer test results, this went to the weight of the evidence and not to its admissibility. See id. at 108-109.