- A. The registrant shall be responsible for directing the operation of the X-ray system(s) under their administrative control. The registrant or the registrant's agent shall assure that the requirements of this Subchapter are met in the operation of the X-ray system(s).
B. The registrant shall have a radiation safety program. The radiation safety program shall include but not be limited to the following:
- 1. The use of ionizing radiation within its purview is performed in accordance with the Act and Agency Regulations.
- 2. All persons are protected as required by Part 1 of this Subchapter.
- C. The registrant shall have a mechanism in place for the referring physician to access information on selecting the most appropriate diagnostic procedure to answer the clinical question.
- D. The registrant shall utilize nationally recognized diagnostic reference levels (DRLs) when applicable.
- E. The registrant shall select the appropriate technique and employ available dose reduction methods and technologies across all patient sizes and clinical indications.
- F. Each registrant shall have a documented procedure in place for verification of patient identity and exam to be performed, including identification of the appropriate body part.
G. All X-ray equipment shall be installed and used in accordance with the equipment manufacturer’s specifications.
4.3.2 Operation Prohibited
An X-ray system which does not meet the provisions of this Part shall not be operated for diagnostic purposes unless the Agency or a Qualified Medical Physicist determines that the non-compliance will not pose a significant radiation risk or significantly affect image quality, and arrangements have been made to correct the non-compliance within thirty (30) days.
4.3.3 Individuals Operating X-ray Systems for Healing Arts Use
- A. Individuals who will be operating the X-ray systems for healing arts use shall possess a current license in accordance with Subchapter 05 Part 34 of this Chapter, Licensure of Radiographers, Nuclear Medicine Technologists, and Radiation Therapists, unless the individual is specifically exempted from licensure by said Regulations. Individuals who will be operating the X-ray systems and who are not subject to licensure under Subchapter 05 Part 34 of this Chapter shall be adequately instructed in the safe operating procedures and be competent in the safe use of the equipment. As a minimum, such instruction shall consist of subjects outlined in § 4.12 of this Part.
- B. The names and qualifications of all personnel operating X-ray equipment for healing arts use must be kept on file for Agency inspection at each facility location.
C. All individuals operating, or supervising the operation of, fluoroscopic X-ray systems shall have completed a minimum of four (4) hours training, prior to performing fluoroscopy procedures that includes but is not limited to the following:
- 1. Biological effects of X-ray;
- 2. Radiation protection methods for patients and staff;
- 3. Factors affecting fluoroscopic outputs;
- 4. Dose management including dose reduction techniques, monitoring, and recording;
- 5. Principles and operation of the specific fluoroscopic X-ray system(s) to be used;
- 6. Fluoroscopic and fluorographic outputs of each mode of operation on the system(s) to be used clinically;
- 7. Basic properties of radiation;
- 8. Units of measurement and dose, including DAP (dose-area product) values and air kerma;
- 9. High level control options; and
- 10. Applicable requirements of this Subchapter.
D. All persons operating, or supervising the operation of, fluoroscopy systems during fluoroscopically-guided interventional (FGI) procedures shall have completed a minimum of eight (8) hours of training approved by the Agency. The training shall include, but not be limited to:
- 1. The topics provided in § 4.3.3(C) of this Part;
- 2. Methods to reduce patient dose using advanced imaging and recording features;
- 3. Procedures for recording pertinent data specified in § 4.5.13 of this Part; and
- 4. A minimum of one (1) hour of hands-on fluoroscopic machine training demonstrating application of topics required in § 4.3.3(D) of this Part.
- E. The training required §§ 4.3.3(C) and (D) of this Part shall be provided by a Qualified Medical Physicist or another individual approved by the Agency.
- F. The registrant shall either provide a minimum of two (2) hours in-service training for all individuals operating or supervising the operation of fluoroscopy systems used, at intervals not to exceed twenty-four (24) months or require evidence of continuing medical education, in fluoroscopic radiation safety and patient dose management at intervals not to exceed twenty-four (24) months.
- G. Documentation pertaining to the requirements of §§ 4.3.3(B), (C) and (D) of this Part shall be maintained for review for three (3) years.
H. A dental facility registrant shall provide initial training and shall ensure that X-ray operators have completed a course in dental radiography offered by an educational institution with a program accredited by the commission on dental accreditation (§ 05-2.10.3(A)(11) of this Chapter). Annual evaluations of X-ray operators to include but not limited to: positioning of the X-ray tube, image processing, operator location during X-ray exposure, source to skin distance, radiation protection, appropriate radiographic protocol, and applicable regulatory requirements. Records of training and annual evaluations shall be maintained for inspection by the Agency for three (3) years.
4.3.4 Technique Protocols
A. For general radiographic systems not equipped with an operational anatomic programming option, protocols shall be documented and readily available to the operator. At a minimum, these protocols shall include the following information:
- 1. Patient's (adult or pediatric, as applicable) body part and anatomical size;
- 2. Technique factors;
- 3. Type of the image receptor to be used;
- 4. Source to image receptor distance to be used (except for dental intraoral radiography, which shall list cone length to be used); and
5. Type of grid, if any.
4.3.5 Written Safety Procedures
The registrant of a facility shall create and make available to X-ray operators written safety procedures, including patient holding and any restrictions of the operating technique required for the safe operation of the particular X-ray system. The operator shall be able to demonstrate familiarity with these procedures.
4.3.6 Room Occupancy During Radiographic Exposure
A. The registrant shall restrict the presence of individuals in the immediate area of the patient being examined, while the X-ray tube is energized, to those required and, as applicable, those in training for the medical procedure, or the parent or guardian of a patient. The following applies to all individuals, other than the patient being examined:
- 1. All individuals shall be positioned such that no part of the body will be struck by the useful beam unless protected by not less than one half millimeter (0.5 mm) lead equivalent material.
- 2. All individuals shall be protected from the secondary radiation by protective garments or whole-body protective barriers of not less than twenty-five one hundredths of one quarter millimeter (0.25 mm) lead equivalent material.
- 3. Instances may warrant having human patients other than the one being examined in the room during the exam. If the procedure results in scatter radiation in excess of two one hundredths (0.02) mSv (2 mR) in any one (1) hour at the position of these patients, such patients shall be protected from the direct scatter radiation by whole body protective barriers of not less than twenty-five one hundredths of one quarter millimeter (0.25 mm) lead equivalent material or shall be positioned so that the two one hundredths (0.02) mSv (2 mR) in any one (1) hour limit is met.
4. Written safety procedures, as required by § 4.3.5 of this Part, shall describe how the requirements of this section will be met when using mobile or portable X-ray systems.
4.3.7 [Reserved]
4.3.8 Non-Healing Arts Exposure Prohibited
Individuals shall not be exposed to the useful beam except for healing arts purposes and unless such exposure has been ordered in writing by a licensed practitioner of the healing arts. This provision specifically prohibits deliberate exposure of an individual for training, demonstration or other non-healing-arts purposes, and exposure of an individual for the purpose of healing arts screening except as authorized by § 4.3.12 of this Part.
4.3.9 When a Patient or Image Receptor Must be Provided with Auxiliary Support During a Radiation Exposure
- A. In cases where a patient or image receptor must be provided with auxiliary support, mechanical support devices shall be used whenever possible. The written safety procedures, required by § 4.3.5 of this Part, shall list individual projections where holding devices cannot be utilized;
- B. Written safety procedures, as required by § 4.3.5 of this Part, shall indicate the requirements for selecting a holder and the procedure the holder shall follow;
- C. The human holder shall be instructed in personal radiation safety and protected as required by § 4.3.6 of this Part;
- D. No individual shall be used routinely to hold image receptor or patients during a radiation exposure; and
E. Each facility shall have protective garments (e.g., aprons, gloves, collars) and shields available in sufficient numbers to provide protection for all personnel who are involved with X-ray operations and who are otherwise not shielded.
- 1. All protective apparel and auxiliary shields shall be evaluated annually for integrity and clearly labeled with their lead equivalence.
F. A record shall be made of the examination and shall include the name of the human holder, date of the examination, number of exposures and technique factors utilized for the exposure(s).
4.3.10 Procedures and Auxiliary Equipment Designed to Minimize Patient and Personnel Exposure Commensurate with The Needed Diagnostic Information Shall Be Utilized
- A. The fastest imaging system consistent with the diagnostic objective of the examinations shall be used. Film cassettes without intensifying screens shall not be used for any diagnostic radiological imaging, with the exception of veterinary radiography and standard film packets for intraoral use in dental radiography.
- B. The radiation exposure to the patient shall be the minimum exposure required to produce images of good diagnostic quality. The registrant shall use auxiliary equipment designed to minimize patient and personnel exposure commensurate with the needed diagnostic information.
- C. Portable or mobile X-ray equipment shall be used only for examinations where it is impractical to transfer the patient(s) to a stationary X-ray installation.
D. X-ray Film Processing Facilities. A registrant using analog image receptors (e.g., radiographic film) shall have available suitable equipment for handling and processing radiographic film in accordance with the following provisions.
1. Manually Developed Film
a. The temperature of solutions in the tanks shall be maintained within the range of sixty degrees Fahrenheit (60º F) to eighty degrees Fahrenheit (80º F) (sixteen degrees Celsius (16º C) to twenty-seven degrees Celsius (27º C)). Film shall be developed in accordance with the time-temperature relationships recommended by the film manufacturer, or, in the absence of such recommendations, with the following time-temperature chart:
| Manual Film Developing Technique Chart |
| Developer Temperature °C/°F | Developing Time (Minutes) | | Developer Temperature °C/°F | Developing Time (Minutes) |
| 26.7/80 | 2.0 | | 20.6/69 | 4.5 |
| 26.1/79 | 2.0 | | 20.0/68 | 5.0 |
| 25.6/78 | 2.5 | | 19.4/67 | 5.5 |
| 25.0/77 | 2.5 | | 18.9/66 | 5.5 |
| 24.4/76 | 3.0 | | 18.3/65 | 6.0 |
| 23.9/75 | 3.0 | | 17.8/64 | 6.5 |
| 23.3/74 | 3.5 | | 17.2/63 | 7.0 |
| 22.8/73 | 3.5 | | 16.7/62 | 8.0 |
| 22.2/72 | 4.0 | | 16.1/61 | 8.5 |
| 21.7/71 | 4.0 | | 15.6/60 | 9.5 |
| 21.1/70 | 4.5 | | | |
b. Devices shall be utilized which will:
- (1) Indicate the actual temperature of the developer; and
- (2) Give an audible or visible signal indicating the termination of a preset time.
- c. Developing tanks shall be constructed of mechanically rigid, corrosion resistant material.
- d. Developing solutions shall be prepared, replenished, and replaced following manufacturer recommendations.
2. Automatic Developers and Other Closed Developing Systems
a. Films shall be developed in accordance with the time temperature relationships recommended by the film manufacturer or, in the absence of such recommendations, with the following time-temperature chart; and
| Developer Temperature | Minimum Immersion Timea/ |
| °C | °F | Seconds |
| 35.5 | 96 | 19 |
| 35 | 95 | 20 |
| 34.5 | 94 | 21 |
| 34 | 93 | 22 |
| 33.5 | 92 | 23 |
| 33 | 91 | 24 |
| 32 | 90 | 25 |
| 31.5 | 89 | 26 |
| 31 | 88 | 27 |
| 30.5 | 87 | 28 |
| 30 | 86 | 29 |
| 29.5 | 85 | 30 |
| a/ Immersion time only, no crossover time included. |
- b. Developing deviations from the requirements of § 4.3.10(D)(2)(a) of this Part shall be documented by the registrant in such manner that the requirements are shown to be met or exceeded (e.g., extended processing, and special rapid chemistry).
- c. Automatic developers shall be operated and maintained following manufacturer specifications.
E. Additional Requirements for Facilities Using Analog Image Receptors
- 1. Pass boxes, if provided, shall be so constructed as to exclude light from the darkroom when cassettes are placed in or removed from the boxes, and shall incorporate adequate shielding from stray radiation to prevent exposure of undeveloped film.
- 2. Facilities shall maintain a light-tight darkroom, use proper safelighting and safeguards, and evaluate darkroom integrity and daylight loading systems for film fog every six (6) months and after any change that may impact film fog.
- 3. Darkrooms typically used by more than one (1) individual shall be provided a method to prevent accidental entry while undeveloped films are being handled or processed.
- 4. Film shall be stored in a cool, dry place and shall be protected from exposure to stray radiation. Film in open packages shall be stored in a light-tight container.
- 5. Film cassettes and intensifying screens shall be inspected periodically and shall be cleaned and replaced as necessary to assure radiographs of good diagnostic quality.
- 6. Outdated X-ray film shall not be used for diagnostic radiographs.
- 7. The film and intensifying screen shall be spectrally compatible.
8. Facilities other than dental, podiatry, and veterinary facilities shall:
- a. Have a continuous and documented sensitometric quality control program, including quality control tests for speed, contrast and fog. These tests shall be performed according to specifications of the manufacturer, a Qualified Medical Physicist, or a nationally recognized organization.
- b. Maintain a light-tight darkroom and use proper safelighting such that any film type in use exposed in a cassette to X-ray radiation sufficient to produce an optical density from one (1) to two (2) when developed shall not suffer an increase in density greater than one tenth (0.1) when exposed in the darkroom for two (2) minutes with all safelights on. If used, daylight film handling systems shall preclude fogging of the film.
- c. Limit the base plus fog of unexposed film to an optical density less than twenty-five one hundredths (0.25) when developed by the routine procedure used by the facility.
F .Facilities Using Computed Radiography (CR) or Direct Digital Radiography (DDR).
- 1. When exposure indicators are available, the facility shall establish and document an acceptable range for the exposure values for examinations routinely performed at the facility. The indicated exposure values for each image shall be compared to the established range. Consistent deviations from established ranges shall be investigated, corrective actions taken as necessary, and results documented.
- 2. CR facilities shall perform at least weekly erasure of all CR cassettes.
G .Dental Facilities
- 1. If using film, the facility shall maintain a light-tight darkroom, use proper safelighting and safeguards, and evaluate darkroom integrity and daylight loading systems for film fog every six (6) months and after any change that may impact film fog.
2. If using a filmless system, the facility shall maintain and operate photostimulable storage phosphor (PSP) and DDR systems according to manufacturer specifications.
4.3.11 Additional Compliance Required
All individuals who are associated with the operation of an X-ray system are subject to the applicable requirements of Parts 1 and 2 of this Subchapter.
4.3.12 Healing Arts Screening
Any person proposing to conduct a healing arts screening program shall not initiate such a program without prior approval of the Agency. When requesting such approval, that person shall submit the information outlined in § 4.11 of this Part. If any information submitted to the Agency becomes invalid or outdated, the Agency shall be immediately notified.
4.3.13 Information and Maintenance Record and Associated Information
A. The registrant shall maintain the following information. The information shall be maintained in a separate file or package in chronological order for each X-ray system, for a minimum of five (5) years or as noted below:
- 1. Maximum rating of technique factors;
- 2. Model and serial numbers of all major components, and user's manuals for those components, including software, shall be maintained for the life of the system;
- 3. Aluminum equivalent filtration in the useful beam, including any routine variation;
- 4. Tube rating charts and cooling curves;
- 5. Records of surveys, calibrations, maintenance, and modifications (e.g., major software and hardware upgrades) performed on the X-ray system(s), with the names of the persons who performed such services;
- 6. A scale drawing of the room in which a stationary X-ray system is located with such drawing indicating the current use of areas adjacent to the room and an estimate of the extent of occupancy by an individual in such areas. In addition, the drawing shall include the results of a survey for radiation levels present at the operator's position, and at pertinent points outside the room at specified test conditions; or the type and thickness of materials, or lead equivalency, of each protective barrier.
7. A copy of all correspondence with this Agency regarding that X-ray system.
4.3.14 X-ray Utilization Log
A. Except for veterinary facilities, each facility shall maintain a record containing the patient's name, the type of examinations, and the dates the examinations were performed. The record shall also include the following information:
- 1. Name of the licensed practitioner of the healing arts ordering the examination.
- 2. Name(s) of individuals who performed the examination.
- 3. Any deviation from the standard procedure as specified on the technique chart, including all repeat exposures.
- 4. When applicable, the X-ray system used.
- 5. When the patient or image receptor must be provided with human auxiliary support, the name of the human holder.
- B. X-ray utilization logs shall be maintained for a minimum of five (5) years following the examination or treatment of adult patients. Records of examination or treatment of minors shall be maintained for a minimum of five (5) years beyond the age of majority.
- C. If X-ray utilization logs are stored electronically, records shall be maintained in a manner that will allow retrieval of records for any specified time period.
D. Each veterinary facility shall maintain a record containing the type of examinations and the dates the examinations were performed. The record shall also include the following information:
- 1. Name(s) of individuals who performed the examination;
- 2. Any deviation from the standard procedure as specified on the technique chart, including all repeat exposures;
- 3. When applicable, the X-ray system used; and
4. When the patient or image receptor must be provided with human auxiliary support, the name of the human holder.
4.3.15 Report and Notification of a Dose to an Embryo/Fetus
- A. A registrant shall report any dose to an embryo/fetus that is greater than fifty (50) mSv (five (5) rem) dose equivalent that is a result of an administration of radiation to a pregnant individual unless the dose to the embryo/fetus was specifically approved, in advance, by the prescribing physician.
- B. The registrant shall notify the Agency by telephone no later than the next business day after discovery of a dose to the embryo/fetus that requires a report in § 4.3.15(A) of this Part.
C. The registrant shall submit a written report, prepared by a Qualified Medical Physicist, to the Agency within fifteen (15) business days after discovery of a dose to the embryo/fetus that requires a report in § 4.3.15(A) of this Part.
1. The written report must include:
- a. The registrant's name and registration number;
- b. The name of the prescribing physician;
- c. A brief description of the event;
- d. Why the event occurred;
- e. The effect, if any, on the embryo/fetus;
- f. What actions, if any, have been taken or are planned to prevent recurrence; and
- g. Certification that the registrant notified the mother (or the mother's or child's responsible relative or guardian), and if not, why not.
- 2. The report must not contain the individual's or child's name or any other information that could lead to identification of the individual or child.
- D. The registrant shall provide notification of the event to the referring physician and also notify the pregnant individual or mother, both hereafter referred to as the mother, no later than twenty-four (24) hours after discovery of an event that would require reporting under § 4.3.15(A) of this Part, unless the referring physician personally informs the registrant either that he or she will inform the mother or that, based on medical judgment, telling the mother would be harmful. The registrant is not required to notify the mother without first consulting with the referring physician. If the referring physician or mother cannot be reached within twenty-four (24) hours, the registrant shall make the appropriate notifications as soon as possible thereafter. The registrant may not delay any appropriate medical care for the embryo/fetus, including any necessary remedial care as a result of the event, because of any delay in notification. To meet the requirements of this paragraph, the notification may be made to the mother's or child's responsible relative or guardian instead of the pregnant individual. If a verbal notification is made, the registrant shall inform the mother, or the mother's or child's responsible relative or guardian, that a written description of the event can be obtained from the registrant upon request. The registrant shall provide such a written description if requested.
E. A registrant shall:
1. Annotate a copy of the report provided to the Agency with the:
- a. Name of the pregnant individual who is the subject of the event; and
- b. Identification number or if no other identification number is available, the Social Security Number of the pregnant individual who is the subject of the event; and
- 2. Provide a copy of the annotated report to the referring physician, if other than the registrant, no later than fifteen (15) days after the discovery of the event.
In addition to other requirements of this Part, all diagnostic and interventional X-ray systems shall meet the requirements of § 4.4 of this Part. Requirements specific to dental intra-oral, panoramic, cephalometric, volumetric dental imaging equipment are included in § 4.14 of this Part.
4.4.2 Maintaining Compliance
Diagnostic and interventional X-ray systems and their associated components used on humans and certified pursuant to the Federal X-ray Equipment Performance Standard (21 C.F.R. Part 1020) shall be maintained in compliance with applicable requirements of that standard.
4.4.3 Warning Label
- A. On systems manufactured on or before June 10, 2006, the control panel containing the main power switch shall bear the following warning statement, or the warning statement in § 4.4.3(B) of this Part, legible and accessible to view: "WARNING: This X-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."
B. On systems manufactured after June 10, 2006, the control panel containing the main power switch shall bear the following warning statement, legible and accessible to view: "WARNING: This X-ray unit may be dangerous to patient and operator unless safe exposure factors, operating instructions and maintenance schedules are observed."
4.4.4 Battery Charge Indicator
On battery-powered X-ray generators, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation.
4.4.5 Leakage Radiation from the Diagnostic Source Assembly
The leakage radiation from the diagnostic source assembly measured at a distance of one meter (1 m) in any direction from the source shall not exceed eighty-eight one hundredths milligray (0.88 mGy) air kerma [one hundred milliroentgen (100 mR) exposure] in one (1) hour when the X-ray tube is operated at its leakage technique factors. If the maximum rated peak tube potential of the tube housing assembly is greater than the maximum rated peak tube potential for the diagnostic source assembly, positive means shall be provided to limit the maximum X-ray tube potential to that of the diagnostic source assembly. Compliance shall be determined by measurements averaged over an area of one hundred square centimeters (100 cm 2) with no linear dimension greater than twenty centimeters (20 cm).
4.4.6 Radiation from Components Other Than the Diagnostic Source Assembly
The radiation emitted by a component other than the diagnostic source assembly shall not exceed an air kerma of eighteen (18) µgray [two (2) milliroentgens exposure] in one (1) hour at five centimeters (5 cm) from any accessible surface of the component when it is operated in an assembled X-ray system under any conditions for which it was designed. Compliance shall be determined by measurements averaged over an area of one hundred square centimeters (100 cm 2) with no linear dimension greater than twenty centimeters (20 cm).
4.4.7 Beam Quality
A. Half-Value Layer (HVL)
- 1. The HVL of the useful beam for a given X-ray tube potential shall not be less than the values shown in § 4.4.7(B) of this Part [Table 1], If it is necessary to determine such half-value layer at an X-ray tube potential which is not listed in § 4.4.7(B) of this Part [Table 1], linear interpolation or extrapolation may be made. Positive means shall be provided to ensure that at least the minimum filtration needed to achieve beam quality requirements is in the useful beam during each exposure. In the case of a system, which is to be operated with more than one (1) thickness of filtration, this requirement can be met by a filter interlocked with the kilovoltage selector which will prevent X-ray emissions if the minimum required filtration is not in place.
- 2. Optional Filtration on Fluoroscopic Systems. Fluoroscopic systems manufactured on or after June 10, 2006, incorporating an X-ray tube(s) with a continuous output of one (1) kilowatt or more and an anode heat storage capacity of one million (1,000,000) heat units or more shall provide the option of adding X-ray filtration to the diagnostic source assembly in addition to the amount needed to meet the half-value layer provisions of § 4.4.7(A)(1) of this Part. The selection of this additional X-ray filtration shall be either at the option of the user or automatic as part of the selected mode of operation. A means of indicating which combination of additional filtration is in the X-ray beam shall be provided.
B. Table 1 – X-Ray Tube Voltage (kilovolt peak)
| Design Operating Range | Measured Operating Potential | Minimum HVL (mm in Aluminum) |
| Subject to § 4.4.7 of this Part and manufactured before June 10, 2006) | X-Ray Systems subject to § 4.4.7 of this Part and manufactured on or after June 10, 2006) |
| Below 51 | 30 | | 0.3 | 0.3 |
| 40 | | 0.4 | 0.4 |
| 50 | | 0.5 | 0.5 |
| 51 to 70 | 51 | | 1.2 | 1.3 |
| 60 | | 1.3 | 1.5 |
| 70 | | 1.5 | 1.8 |
| Above 70 | 71 | | 2.1 | 2.5 |
| 80 | | 2.3 | 2.9 |
| 90 | | 2.5 | 3.2 |
| 100 | | 2.7 | 3.6 |
| 110 | | 3.0 | 3.9 |
| 120 | | 3.2 | 4.3 |
| 130 | | 3.5 | 4.7 |
| 140 | | 3.8 | 5.0 |
| 150 | | 4.1 | 5.4 |
- C. Measuring Compliance. For capacitor energy storage equipment, compliance shall be determined with the maximum selectable quantity of charge per exposure.
- D. Aluminum Equivalent of Material Between Patient and Image Receptor. Except when used in a CT X-ray system, the aluminum equivalent of each of the items listed in § 4.4.7(E) of this Part [Table 2], which are used between the patient and the image receptor, may not exceed the indicated limits. Compliance shall be determined by X-ray measurements made at a potential of one hundred (100) kilovolts peak and with an X-ray beam that has an HVL specified in § 4.4.7(B) of this Part [Table 1] for the potential. This requirement applies to front panel(s) of cassette holders and film changers provided by the manufacturer for patient support or for prevention of foreign object intrusions. It does not apply to screens and their associated mechanical support panels or grids.
E. Table 2 – Maximum Aluminum Equivalent (millimeters)
| ITEM | Maximum Aluminum Equivalent (millimeters) |
| 1. Front panel(s) of cassette holders (total of all) | 1.2 |
| 2. Film panel(s) of film changer (total of all) | 1.2 |
| 3. Cradle | 2.3 |
| 4. Tabletop, stationary, without articulated joints | 1.2 |
| 5. Tabletop, movable, without articulated joint(s) (including stationary subtop) | 1.7 |
| 6. Tabletop, with radiolucent panel having one articulated joint | 1.7 |
| 7. Tabletop, with radiolucent panel having two or more articulated joints | 2.3 |
| 8. Tabletop, cantilevered | 2.3 |
| 9. Tabletop, radiation therapy simulator | 5.0 |
F. Modification of Certified Diagnostic X-ray Components and Systems
- 1. Diagnostic X-ray components and systems certified in accordance with 21 C.F.R. Part 1020 shall not be modified such that the component or system fails to comply with any applicable provision of this Part.
2. The owner of a diagnostic X-ray system who uses the system in a professional or commercial capacity may modify the system provided the modification does not result in the failure of the system or component to comply with the applicable requirements of this Part. The owner who causes such modification need not submit the reports required by this Subchapter, provided the owner records the date and the details of the modification in the system records and maintains this information, and provided the modification of the X-ray system does not result in a failure to comply with this Subchapter.
4.4.8 Multiple Tubes
Where two (2) or more radiographic tubes are controlled by one (1) exposure switch, the tube or tubes which have been selected shall be clearly indicated prior to initiation of the exposure. This indication shall be both on the X-ray control panel and at or near the tube housing assembly which has been selected.
4.4.9 Mechanical Support of Tube Head
The tube housing assembly supports shall be adjusted such that the tube housing assembly will remain stable during an exposure unless tube housing movement is a designed function of the X-ray system.
4.4.10 Technique Indicators
- A. For X-ray equipment capable of displaying technique factors, the technique factors to be used during an exposure shall be indicated before the exposure begins. If automatic exposure controls are used, the technique factors which are set prior to the exposure shall be indicated.
- B. The requirement of § 4.4.10(A) of this Part may be met by permanent markings on equipment having fixed technique factors. Indication of technique factors shall be visible from the operator's position except in the case of spot films made by the fluoroscopist.
C. The accuracy of the indicated kilovoltage peak (kVp) shall meet manufacturer specifications. In the absence of a manufacturer specification, kVp accuracy shall be within ± ten percent (10%).
4.4.11 Structural Shielding
Structural shielding shall be provided whenever necessary to meet the requirements of §§ 1.7.1 and 1.8.1 of this Subchapter, in addition to specific requirements contained in other Parts of this Subchapter.
4.4.12 Locks
All position locking, holding, and centering devices on X-ray system components and systems shall function as intended.
4.4.13 Use of Calibrated Dosimetry System
Measurements required in §§ 4.10.1(A)(3), (4) and (5) of this Part shall be performed with a calibrated dosimetry system per manufacturer recommendations. The calibration of such a system shall be traceable to a national standard. The dosimetry system shall have been calibrated within the preceding two (2) years and records of calibration maintained for five (5) years for inspection by the Agency.
4.4.14 Reports and Notifications of Radiation Medical Events
A. Other than events that result from intervention by a patient or human research subject, a registrant shall report any event in which the administration of ionizing radiation from a diagnostic radiation machine meets one (1) or more of the following criteria:
- 1. A patient or human research subject receives an unintended skin dose to the same area in a single procedure greater than two (2) Gy [two hundred (200) rads].
2. A patient or human research subject receives an unintended dose other than skin dose in a single procedure greater than:
- a. Five (5) times the facility’s established protocol, and > 0.5 Gy (50 rad) to any organ; or
- b. Five (5) times the facility’s established protocol, and > 0.05 Sv (5 rem) total effective dose.
3. Wrong patient or wrong site for the entire procedure when the resultant dose:
- a. > 0.5 Gy (50 rad) to any organ; or
- b. Total Effective dose is greater than or equal to ( =) 0.05 Sv (5 rem).
- B. Any wrong patient or wrong site imaged regardless of dose received shall be reported, documented and addressed internally within the facility.
C. The registrant shall notify the Agency by telephone no later than the next business day after discovery of the radiation medical event.
- 1. All required notifications shall use Agency contact information specified in § 1.4 of this Subchapter.
D. The registrant shall submit a written report to the Agency, prepared by a Qualified Medical Physicist, to the Agency within fifteen (15) business days after discovery of the radiation medical event. The written report shall include:
- 1. The registrant’s name;
- 2. Date of event and date discovered;
- 3. The estimated dose received;
- 4. The imaging procedure(s) performed;
- 5. The type of equipment in use (e.g., CT, fluoroscopy, radiographic, other);
- 6. The manufacturer and model of the unit used;
- 7. Why the event occurred;
- 8. How the event was discovered;
- 9. The effect, if any, on the individuals(s) who is the subject of the radiation medical event;
- 10. Actions, if any, that have been taken, or are planned, to prevent recurrence;
- 11. Certification that the registrant notified the individual (or the individual’s responsible relative or guardian), and if not, why not; and
- 12. If there was notification, what information was provided to the individual.
E. The registrant shall provide a clinical summary of the radiation medical event to the prescribing physician and patient within fifteen (15) business days.
4.4.15 Records of Radiation Medical Events
- The registrant shall retain a record of a radiation medical event reported in accordance with § 4.4.14 of this Part as part of the patient's permanent medical record.
- A. The provisions of § 4.5 of this Part apply to equipment for fluoroscopic imaging or for recording images from the fluoroscopic image receptor. X-ray systems subject to § 4.5 of this Part shall also meet the requirements of § 4.4 of this Part.
B. Only image-intensified or direct-digital receptor fluoroscopic equipment shall be used for fluoroscopy.
4.5.2 Primary Protective Barrier
- A. Limitation of Useful Beam. The fluoroscopic imaging assembly shall be provided with a primary protective barrier which intercepts the entire cross section of the useful beam at any SID. The X-ray tube used for fluoroscopy shall not produce X-rays unless the barrier is in position to intercept the entire useful beam. The air kerma rate [AKR] due to transmission through the barrier with the attenuation block in the useful beam combined with radiation from the fluoroscopic imaging receptor shall not exceed 3.34 x 10-3 percent of the entrance AKR, at a distance of ten centimeters (10 cm) from any accessible surface of the fluoroscopic imaging assembly beyond the plane of the image receptor.
B. Measuring Compliance. The AKR shall be measured in accordance with § 4.5.6 of this Part. The AKR due to transmission through the primary barrier combined with radiation from the fluoroscopic image receptor shall be determined by measurements averaged over an area of one hundred square centimeters (100 cm2) with no linear dimension greater than twenty centimeters (20 cm). If the source is below the tabletop, the measurement shall be made with the input surface of the fluoroscopic imaging assembly positioned thirty centimeters (30 cm) above the tabletop. If the source is above the tabletop and the SID is variable, the measurement shall be made with the end of the beam-limiting device or spacer as close to the tabletop as it can be placed, provided that it shall not be closer than thirty centimeters (30 cm). Movable grids and compression devices shall be removed from the useful beam during the measurement. For all measurements, the attenuation block shall be positioned in the useful beam ten centimeters (10 cm) from the point of measurement of entrance AKR and between this point and the input surface of the fluoroscopic imaging assembly.
4.5.3 Equipment Operation
- A. All fluoroscopic images shall be viewed, directly or indirectly, and interpreted by a licensed practitioner of the healing arts.
- B. The operation of mobile or portable fluoroscopic X-ray systems, by radiologic technologists, for positioning purposes only, shall be performed under the direct supervision of a licensed practitioner of the healing arts who meets the requirements of § 4.3.3(C) of this Part.
- C. A medical resident or radiologic technology student in training shall not be allowed to operate fluoroscopic X-ray systems unless in the physical presence of a licensed practitioner of the healing arts and a radiologic technologist, as specified in § 4.3.3(C) of this Part.
- D. Overhead fluoroscopy shall not be used as a positioning tool for general purpose radiographic examinations.
- E. Operators shall be competent in the standard operating procedures of the unit in use, including the use of available dose-saving features, and the relative radiation output rates of the various modes of operation.
- F. Procedure planning for fluoroscopic procedures on pregnant patients shall include feasible modifications to minimize the dose to the conceptus.
- G. Procedure planning for fluoroscopic procedures on pediatric patients shall include feasible modifications to minimize dose.
- H. The registrant shall use all methods available on the fluoroscopy system to monitor dose during a fluoroscopic procedure.
I. The facility shall establish a written policy regarding patient dose management in fluoroscopically guided procedures in conformance with the ACR-AAPM Technical Standard for Management of the Use of Radiation in Fluoroscopic Procedures (2021) incorporated above at § 4.1.1(B) of this Part or NCRP Report 168 Radiation Dose Management for Fluoroscopically-Guided Interventional Medical Procedures incorporated above at § 4.1.1(C) of this Part.
4.5.4 Field Limitation
- A. Angulation. For fluoroscopic equipment manufactured after February 25, 1978, when the angle between the image receptor and the beam axis of the X-ray beam is variable, means shall be provided to indicate when the axis of the X-ray beam is perpendicular to the plane of the image receptor. Compliance with §§ 4.5.4(C) and (D) of this Part shall be determined with the beam axis indicated to be perpendicular to the plane of the image receptor.
- B. Further Means for Limitation. Means shall be provided to permit further limitation of the X-ray field to sizes smaller than the limits of §§ 4.5.4(C) and (D) of this Part. Beam-limiting devices manufactured after May 22,1979 and incorporated in equipment with a variable SID and/or capability of a visible area of greater than three hundred square centimeters (300 cm2), shall be provided with means for stepless adjustment of the X-ray field. Equipment with a fixed SID and the capability of a visible area of no greater than three hundred square centimeters (300 cm2) shall be provided with either stepless adjustment of the X-ray field or with a means to further limit the X-ray field size at the plane of the image receptor to one hundred twenty-five square centimeters (125 cm2) or less. Stepless adjustment shall, at the greatest SID, provide continuous field sizes from the maximum obtainable to a field size containable in a square of five centimeters (5 cm) by five centimeters (5 cm).
C. Fluoroscopy and Radiography Using the Fluoroscopic Imaging Assembly with Inherently Circular Image Receptors
1. For fluoroscopic equipment manufactured before June 10, 2006, other than radiation therapy simulation systems, the following applies:
- a. Neither the length nor width of the X-ray field in the plane of the image receptor shall exceed that of the visible area of the image receptor by more than three percent (3%) of the SID. The sum of the excess length and the excess width shall be no greater than four percent (4%) of the SID.
- b. For rectangular X-ray fields used with circular image receptors, the error in alignment shall be determined along the length and width dimensions of the X-ray field which pass through the center of the visible area of the image receptor.
2. For fluoroscopic equipment manufactured on or after June 10, 2006, other than radiation therapy simulation systems, the maximum area of the X-ray field in the plane of the image receptor shall conform with one (1) of the following requirements:
- a. When any linear dimension of the visible area of the image receptor measured through the center of the visible area is less than or equal to thirty-four centimeters (34 cm) in any direction, at least eighty percent (80%) of the area of the X-ray field overlaps the visible area of the image receptor, or
- b. When any linear dimension of the visible area of the image receptor measured through the center of the visible area is greater than thirty-four centimeters (34 cm) in any direction, the X-ray field measured along the direction of greatest misalignment with the visible area of the image receptor does not extend beyond the edge of the visible area of the image receptor by more than two centimeters (2 cm).
D. Fluoroscopy and Radiography Using Fluoroscopic Imaging Assembly with Inherently Rectangular Image Receptors. For X-ray systems manufactured on or after June 10, 2006, the following applies:
- 1. Neither the length nor width of the X-ray field in the plane of the image receptor shall exceed that of the visible area of the image receptor by more than three percent (3%) of the SID. The sum of the excess length and the excess width shall be no greater than four percent (4%) of the SID.
- 2. The error in alignment shall be determined along the length and width dimensions of the X-ray field which pass through the center of the visible area of the image receptor.
E. Override Capability. If the fluoroscopic X-ray field size is adjusted automatically as the SID or image receptor size is changed, a capability may be provided for overriding the automatic adjustment in case of system failure. If it is so provided, a signal visible at the operator’s position shall indicate whenever the automatic field adjustment is overridden. Each such system failure override switch shall be clearly labeled as follows: FOR X-RAY FIELD LIMITATION SYSTEM FAILURE.
4.5.5 Activation of the Tube
X-ray production in the fluoroscopic mode shall be controlled by a device which requires continuous pressure by the operator for the entire time of any exposure. When recording serial fluoroscopic images from the fluoroscopic image receptor, the operator shall be able to terminate the X-ray exposure(s) at any time but means may be provided to permit completion of any single exposure of the series in process.
4.5.6 Air Kerma Rates (AKR)
A. For fluoroscopic equipment, the following requirements apply:
1. Fluoroscopic equipment manufactured before May 19, 1995.
- a. Equipment provided with automatic exposure rate control (AERC) shall not be operable at any combination of tube potential and current that will result in an AKR in excess of eighty-eight (88) mGy per minute (ten (10) R/min) exposure rate at the measurement point specified in 21 C.F.R. § 1020.32(d)(3), except as specified in § 4.5.6(A)(1)(e) of this Part.
- b. Equipment provided without AERC shall not be operable at any combination of tube potential and current that will result in an AKR in excess of forty-four (44) mGy per minute (five (5) R/min exposure rate) at the measurement point specified in 21 C.F.R. § 1020.32(d)(3), except as specified in § 4.5.6(A)(1)(e) of this Part.
- c. Equipment provided with both an AERC mode and a manual mode shall not be operable at any combination of tube potential and current that will result in an AKR in excess of eighty-eight (88) mGy per minute (ten (10) R/min exposure rate) in either mode at the measurement point specified in 21 C.F.R. § 1020.32(d)(3), except as specified in § 4.5.6(A)(1)(e) of this Part.
- d. Equipment may be modified in accordance with § 4.4.7(F) of this Part to comply with § 4.5.6(A)(2) of this Part. When the equipment is modified, it shall bear a label indicating the date of the modification and the statement: MODIFIED TO COMPLY WITH 21 C.F.R. § 1020.32(H)(2).
e. Exceptions:
- (1) During recording of fluoroscopic images.
2. Fluoroscopic equipment manufactured on or after May 19, 1995
- a. Shall be equipped with AERC if operable at any combination of tube potential and current that results in an AKR greater than forty-four (44) mGy per minute (five (5) R/min exposure rate) at the measurement point specified in 21 C.F.R. § 1020.32(d)(3). Provision for manual selection of technique factors may be provided.
- b. Shall not be operable at any combination of tube potential and current that will result in an AKR in excess of eighty-eight (88) mGy per minute (ten (10) R/min exposure rate) at the measurement point specified in 21 C.F.R. § 1020.32(d)(3), except as specified in § 4.5.6(A)(2)(c) of this Part.
c. Exceptions
- (1) For equipment manufactured prior to June 10, 2006, during the recording of images from a fluoroscopic image receptor using photographic film or a video camera when the X-ray source is operated in a pulsed mode.
- (2) For equipment manufactured on or after June 10, 2006, during the recording of images from the fluoroscopic image receptor for the purpose of providing the user with a recorded image(s) after termination of the exposure. Such recording does not include images resulting from a last-image-hold feature that are not recorded.
(3) Fluoroscopy equipment with optional high-level control. When high-level control is selected and the control is activated, in which case the equipment shall not be operable at any combination of tube potential and current that will result in an AKR in excess of one hundred seventy-six (176) mGy per minute (twenty (20) R/min exposure rate) at the measurement point specified in 21 C.F.R. § 1020.32(d)(3). Special means of activation of high-level controls shall be required. The high-level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the operator shall indicate that the high-level control is employed.
4.5.7 Measurement of Entrance AKR
A. Compliance with § 4.5.6 of this Part shall be determined as follows:
- 1. If the source is below the X-ray table, the AKR shall be measured at one centimeter (1 cm) above the tabletop or cradle.
- 2. If the source is above the X-ray table, the AKR shall be measured at thirty centimeters (30 cm) above the tabletop with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement.
- 3. In a C-arm type of fluoroscope, the AKR shall be measured at thirty centimeters (30 cm) from the input surface of the fluoroscopic imaging assembly, with the source positioned at any available SID, provided that the end of the beam-limiting device or spacer is no closer than thirty centimeters (30 cm) from the input surface of the fluoroscopic imaging assembly.
- 4. In a C-arm type of fluoroscope having a SID less than forty-five centimeters (45 cm), the AKR shall be measured at the minimum SSD.
5. In a lateral type of fluoroscope, the AKR shall be measured at a point fifteen centimeters (15 cm) from the centerline of the X-ray table and in the direction of the X-ray source with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement. If the tabletop is movable, it shall be positioned as closely as possible to the lateral X-ray source, with the end of the beam-limiting device or spacer no closer than fifteen centimeters (15 cm) to the centerline of the X-ray table.
4.5.8 Indication of Potential and Current
During fluoroscopy and cinefluorography, the X-ray tube potential and current shall be continuously indicated. Deviation of X-ray tube potential and current from the indicated value shall not exceed the maximum deviation as stated by the manufacturer.
4.5.9 Source-Skin Distance
- A. Means shall be provided to limit the source-skin distance to not less than thirty-eight centimeters (38 cm) on stationary fluoroscopes and to not less than thirty centimeters (30 cm) on mobile and portable fluoroscopes. In addition, for fluoroscopes intended for specific surgical application that would be prohibited at the source-skin distances specified in this paragraph, provisions may be made for operating at shorter source-skin distances but in no case less than twenty centimeters (20 cm).
B. For stationary, mobile, or portable C-arm fluoroscopic systems manufactured on or after June 10, 2006, having a maximum source-image receptor distance of less than forty-five centimeters (45 cm), means shall be provided to limit the source-skin distance to not less than nineteen centimeters (19 cm). Such systems shall be labeled for extremity use only. In addition, for those systems intended for specific surgical application that would be prohibited at the source-skin distance specified in this paragraph, provisions may be made for operation at shorter source-skin distances but in no case less than ten centimeters (10 cm).
4.5.10 Fluoroscopic Irradiation Time, Display and Signal
A. Fluoroscopic equipment manufactured before June 10, 2006:
- 1. Shall be provided with means to preset the cumulative irradiation time of the fluoroscopic tube. The maximum cumulative time of the timing device shall not exceed five (5) minutes without resetting. A signal audible to the fluoroscopist shall indicate the completion of any preset cumulative irradiation time. Such signal shall continue to sound while X-rays are produced until the timing device is reset. Fluoroscopic equipment may be modified in accordance with 21 C.F.R. § 1020.30(q) to comply with the requirements of § 4.5.10 of this Part. When the equipment is modified, it shall bear a label indicating the statement: MODIFIED TO COMPLY WITH 21 C.F.R. § 1020.32(h)(2).
B. For X-ray controls manufactured on or after June 10, 2006, there shall be provided for each fluoroscopic tube:
1. A display of the fluoroscopic irradiation time at the fluoroscopist’s working position. This display shall function independently of the audible signal described § 4.5.10(B)(2) of this Part. The following requirements apply:
- a. When the X-ray tube is activated, the fluoroscopic irradiation time in minutes and tenths of minutes shall be continuously displayed and updated at least once every six (6) seconds.
- b. The fluoroscopic irradiation time shall also be displayed within six (6) seconds of termination of an exposure and remain displayed until reset.
- c. Means shall be provided to reset the display to zero (0) prior to the beginning of a new examination or procedure.
2. A signal audible to the fluoroscopist shall sound for each passage of five (5) minutes of fluoroscopic irradiation time during an examination or procedure. The signal shall sound until manually reset or, if automatically reset, for at least two (2) seconds.
4.5.11 [RESERVED]
4.5.12 Protection From Scattered Radiation
- A. For stationary fluoroscopic systems, ancillary shielding, such as drapes, self-supporting curtains, or viewing shields, shall be available and used as supplemental protection for all individuals other than the patient in the room during a fluoroscopy procedure.
B. Where sterile fields or special procedures prohibit the use of normal protective barriers or drapes, all of the following conditions shall be met:
- 1. Shielding required under § 4.5.12(A) of this Part shall be maintained to the degree possible under the clinical conditions;
- 2. All persons, except the patient, in the room where fluoroscopy is performed shall wear protective garments that provide a lead equivalent shielding of at least twenty-five one hundredths of one millimeter (0.25 mm);
- 3. The fluoroscopic field size shall be reduced to the minimum required for the procedure being performed (area of clinical interest); and
4. Operating and safety procedures shall reflect the above conditions, and fluoroscopy personnel shall exhibit awareness of situations requiring the use and/or non-use of the protective drapes.
4.5.13 Additional Requirements for Facilities Performing Fluoroscopically-Guided Interventional (FGI) Procedures
A. A registrant utilizing fluoroscopically-guided interventional (FGI) procedures shall establish a Radiation Protocol Committee (RPC) in accordance with the following:
- 1. The registrant may establish a system-wide committee if the registrant has more than one (1) site;
- 2. Two (2) or more registrants may form a cooperative RPC as long as each facility has a representative on the committee; and
- 3. If the registrant has already established a radiation safety committee, the requirements of § 4.5.13(A) of this Part may be delegated to that committee if the members meet the requirements of § 4.5.13(E) of this Part.
- B. A quorum of the RPC shall meet as often as necessary, but at intervals not to exceed twelve (12) months.
- C. A record of each RPC meeting shall include the date, names of individuals in attendance, minutes of the meeting, and any actions taken. The registrant shall maintain the record for inspection by the Agency for three (3) years.
- D. The RPC shall provide an annual report to the radiation safety committee or radiation safety officer, in the absence of a radiation safety committee.
E. Members of the RPC shall include but not be limited to the following individuals:
- 1. A supervising physician of the healing arts who meets the requirements in § 4.3.3(C) of this Part;
- 2. A Qualified Medical Physicist;
- 3. The lead technologist; and
- 4. Other individuals as deemed necessary by the registrant.
F. The RPC shall establish and implement written FGI procedure protocols, or protocols documented in an electronic report system, that include but are not limited to the following:
- 1. Identification of individuals who are authorized to use fluoroscopic systems for interventional purposes;
- 2. A method to be used to monitor patient radiation dose during FGI;
- 3. Dose notification levels, as appropriate, at which the physician is notified, and appropriate actions are taken for patient safety;
- 4. Substantial radiation dose level (SRDL) values following nationally recognized standards;
- 5. Actions to be taken for cases when a SRDL is exceeded which may include patient follow-up; and
- 6. A review of the established protocols at an interval not to exceed twelve (12) months.
- G. A record of each RPC protocol shall be maintained for inspection by the Agency. If the RPC revises a protocol, documentation shall be maintained that includes the justification for the revision and the previous protocol for inspection by the Agency.
H. A record of radiation output information shall be maintained, for three (3) years for inspection by the Agency so the radiation dose to the skin may be estimated in accordance with established protocols. The record shall include, but not be limited to, the following:
- 1. Patient identification;
- 2. Type and date of examination;
- 3. Identification of the fluoroscopic system used; and
4. Peak skin dose, cumulative air kerma or dose area product used if the information is available on the fluoroscopic system. If the peak skin dose, cumulative air kerma or dose area product are not displayed on the fluoroscopic system, records shall include other information necessary to estimate the radiation dose to the skin in accordance with established protocol or the following as necessary:
- a. Fluoroscopic mode, such as, high-level or pulsed mode of operation;
- b. Cumulative fluoroscopic exposure time; and
c. Number of films or recorded exposures.
4.5.14 Radiation Therapy Simulation Systems
- A. Radiation therapy simulation systems shall be exempt from the requirements of § 4.5.2(A) of this Part, provided such systems are intended only for remote control operation.
- B. Radiation therapy simulation systems shall be exempt from all the requirements of §§ 4.5.4(C), 4.5.6, and 4.6.12(A)(2) of this Part when used for therapy simulation purposes.
C. As an alternative to the requirements of § 4.5.10 of this Part, radiation therapy simulation systems may be provided with a means to indicate the total cumulative exposure time during which X-rays were produced, and which is capable of being reset between X-ray examinations.
4.5.15 Display of Last-Image-Hold (LIH)
- A. Fluoroscopic equipment manufactured on or after June 10, 2006, shall be equipped with means to display LIH image following termination of the fluoroscopic exposure.
- B. For an LIH image obtained by retaining pretermination fluoroscopic images, if the number of images and method of combining images are selectable by the user, the selection shall be indicated prior to initiation of the fluoroscopic exposure.
- C. For an LIH image obtained by initiating a separate radiographic-like exposure at the termination of fluoroscopic imaging, the technique factors for the LIH image shall be selectable prior to the fluoroscopic exposure, and the combination selected shall be indicated prior to initiation of the fluoroscopic exposure.
D. Means shall be provided to clearly indicate to the user whether a displayed image is the LIH radiograph or fluoroscopy. Display of the LIH radiograph shall be replaced by the fluoroscopic image concurrently with re-initiation of fluoroscopic exposure, unless separate displays are provided for the LIH radiograph and fluoroscopic images.
4.5.16 Displays of Values of Air Kerma Rate (AKR) and Cumulative Air Kerma
A. Fluoroscopic equipment manufactured on or after June 10, 2006, shall display at the fluoroscopist’s working position the AKR and cumulative air kerma. The following requirements apply for each X-ray tube used during an examination or procedure:
- 1. When the X-ray tube is activated and the number of images produced per unit time is greater than six (6) images per second, the AKR in mGy/min shall be continuously displayed and updated at least once every second.
- 2. The cumulative air kerma in units of mGy shall be displayed either within five (5) seconds of termination of an exposure or displayed continuously and updated at least once every five (5) seconds.
- 3. The display of the AKR shall be clearly distinguishable from the display of the cumulative air kerma.
4. The AKR and cumulative air kerma shall represent the value for conditions of free-in-air irradiation at one of the following reference locations specified according to the type of fluoroscope.
- a. For fluoroscopes with X-ray source below the X-ray table, X-ray source above the table, or of lateral type, the reference location shall be the respective locations specified in §§ 4.5.7(A)(1), (2) or (5) of this Part.
- b. For C-arm fluoroscopes, the reference location shall be fifteen (15) cm from the isocenter toward the X-ray source along the beam axis. Alternatively, the reference location shall be at a point specified by the manufacturer to represent the location of the intersection of the X-ray beam with the patient’s skin.
- 5. Means shall be provided to reset to zero (0) the display of cumulative air kerma prior to the commencement of a new examination or procedure.
- 6. The displayed AKR and cumulative air kerma shall not deviate from the actual values by more than ± thirty-five percent (± 35%) over the range of six (6) mGy/min and one hundred (100) mGy to the maximum indication of AKR and cumulative air kerma, respectively. Compliance shall be determined with an irradiation time greater than three (3) seconds.
The provisions of § 4.6 of this Part apply to all non-dental registrants using diagnostic X-ray equipment. X-ray systems subject to § 4.6 of this Part shall also meet the requirements of § 4.4 of this Part. Requirements specific to using dental intra-oral, hand held, panoramic, and cephalometric equipment are in §§ 4.13 and 4.14 of this Part. The useful beam shall be limited to the area of clinical interest.
4.6.2 Radiation Exposure Control
- A. Exposure Initiation. Means shall be provided to initiate the radiation exposure by a deliberate action on the part of the operator, such as the depression of a switch. Radiation exposure shall not be initiated without such an action. In addition, it shall not be possible to initiate an exposure when the timer is set to a "zero" or "off" position if either position is provided.
- B. Exposure Indication. Means shall be provided for visual indication observable at or from the operator's protected position whenever X-rays are produced. In addition, a signal audible to the operator shall indicate that the exposure has terminated.
C. Operator Protection, Except Veterinary Systems
- 1. Stationary Radiographic Systems. Stationary radiographic systems shall be required to have the X-ray control, including the exposure switch, permanently mounted in a protected area so that the operator is required to remain in that protected area during the entire exposure.
2. Mobile and Portable Systems. Mobile and portable X-ray systems which are:
- a. Used continuously for greater than one (1) week in the same location (i.e., a room or suite) shall meet the requirements of § 4.6.2(C)(1) of this Part;
- b. Used for less than one (1) week at the same location shall be provided with either a protective barrier at least two meters (2 m) (six and one half feet (6.5’)) high for operator protection during exposures or means shall be provided to allow the operator to be at least two and seven tenths meters (2.7 m) (nine feet (9’)) from the tube housing assembly during the exposure.
- 3. Podiatry Systems. Podiatry facilities shall meet the protection requirements in § 4.6.2(C)(2)(b) of this Part.
D. Operator and Ancillary Personnel Protection for Veterinary Systems
1. All stationary, mobile or portable X-ray systems used for veterinary work shall be provided with either a two meter (2 m) (six and one half feet (6.5’)) high protective barrier for operator protection during exposures or shall be provided with means to allow the operator to be at least two and seven tenths meters (2.7 m) (nine feet (9’)) from the tube housing assembly during exposures. Otherwise, in cases where animals are held, the operator and ancillary personnel shall be protected by a minimum of twenty-five one hundredths millimeter (0.25 mm) lead equivalent from scatter radiation and one half millimeter (0.5 mm) from the useful beam. Refer to § 4.13 of this Part for hand-held intraoral dental radiographic units used in veterinary practice.
4.6.3 Control and Indication of Technique Factors
A. Timers. Means shall be provided to terminate the exposure at a preset time interval, a preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor.
- 1. Except during serial radiography, the operator shall be able to terminate the exposure at any time during an exposure of greater than one half (0.5) second. Except during panoramic dental radiography, termination of exposure shall cause automatic resetting of the timer to its initial setting or to zero (0). It shall not be possible to make an exposure when the timer is set to a zero (0) or off position if either position is provided.
- 2. During serial radiography, the operator shall be able to terminate the X-ray exposure(s) at any time but means may be provided to permit completion of any single exposure of the series in process.
B. Automatic Exposure Controls. When an automatic exposure control is provided:
- 1. Indication shall be made on the control panel when this mode of operation is selected;
- 2. When the X-ray tube potential is equal to or greater than fifty-one kilovolts peak (51 kVp), the minimum exposure time for field emission equipment rated for pulse operation shall be equal to or less than a time interval equivalent to two pulses and the minimum exposure time for all other equipment shall be equal to or less than one sixtieth of one (1/60) second or a time interval required to deliver five milliampere-seconds (5 mAs), whichever is greater;
- 3. Either the product of peak X-ray tube potential, current, and exposure time shall be limited to not more than sixty kilowatt-seconds (60 kWs) per exposure or the product of X-ray tube current and exposure time shall be limited to not more than six hundred (600) mAs per exposure, except when the X-ray tube potential is less than fifty-one (51) kVp, in which case the product of X-ray tube current and exposure time shall be limited to not more than two thousand (2,000) mAs per exposure; and
- 4. A visible signal shall indicate when an exposure has been terminated at the limits described in § 4.6.3(B)(3) of this Part, and manual resetting shall be required before further automatically timed exposures can be made.
C. Accuracy. Deviation of technique factors under § 4.6.3 of this Part from indicated values shall not exceed the limits given by the manufacturer.
4.6.4 Positive Beam Limitation (PBL)
- A. The requirements of § 4.6.4 of this Part shall apply to radiographic systems which contain PBL.
B. Field Size. When a PBL system is provided, it shall prevent X-ray production when:
- 1. Either the length or width of the X-ray field in the plane of the image receptor differs from the corresponding image receptor dimension by more than three percent (3%) of the SID; or
- 2. The sum of the length and width differences stated in § 4.6.4(B)(1) of this Part without regard to sign exceeds four percent (4%) of the SID.
- 3. The beam-limiting device is at a SID for which PBL is not designed for sizing.
C. Conditions For PBL. When provided, the PBL system shall function as described in § 4.6.4(B) of this Part whenever all the following conditions are met:
- 1. The image receptor is inserted into a permanently mounted cassette holder;
- 2. The image receptor length and width are less than fifty centimeters (50 cm);
- 3. The X-ray beam axis is within ± three degrees (± 3°) of vertical and the SID is ninety centimeters (90 cm) to one hundred thirty centimeters (130 cm) inclusive; or the X-ray beam axis is within ± three degrees (± 3°) of horizontal and the SID is ninety centimeters (90 cm) to two hundred five centimeters (205 cm) inclusive;
- 4. The X-ray beam axis is perpendicular to the plane of the image receptor to within ± three degrees (± 3°); and
- 5. Neither tomographic nor stereoscopic radiography is being performed.
- D. Measuring Compliance. Compliance with the requirements of § 4.6.4(B) of this Part shall be determined when the equipment indicates that the beam axis is perpendicular to the plane of the image receptor and the provisions of § 4.6.4(B) of this Part are met. Compliance shall be determined no sooner than five (5) seconds after insertion of the image receptor.
- E. Operator Initiated Undersizing. The PBL system shall be capable of operating such that, at the discretion of the operator, the size of the field may be made smaller than the size of the image receptor through stepless adjustment of the field size. Each dimension of the minimum field size at a SID of one hundred centimeters (100 cm) shall be equal to or less than five centimeters (5 cm). Return to PBL function as described in § 4.6.4(B) of this Part shall occur automatically upon any change of image receptor size or SID.
F. Override of PBL. A capability may be provided for overriding PBL in case of system failure and for servicing the system. This override may be for all SIDs and image receptor sizes. A key shall be required for any override capability that is accessible to the operator. It shall not be possible to remove the key while PBL is overridden. Each such key switch or key shall be clearly and durably labeled as follows: FOR X-RAY FIELD LIMITATION SYSTEM FAILURE
- 1. The override capability is considered accessible to the operator if it is referenced in the operator’s manual or in other material intended for the operator or if its location is such that the operator would consider it part of the operational controls.
G. Disabling of PBL. A facility has the option to permanently functionally disable a PBL system. When this option is chosen, the standards for manual collimation apply.
4.6.5 Source-to-Skin Distance
The minimum source-skin distance shall not be less than thirty centimeters (30 cm), except intraoral dental equipment covered under § 4.14 of this Part and veterinary equipment.
4.6.6 Air Kerma Reproducibility
A. Coefficient of variation. For any specific combination of selected technique factors, the estimated coefficient of variation of the air kerma shall be no greater than five one hundredths (0.05).
- 1. Measuring compliance. Determination of compliance shall be based on five (5) consecutive measurements taken within a time period of one (1) hour. An additional five (5) consecutive measurements may be taken (ten (10) consecutive measurements total) to determine compliance if the results of the initial five (5) measurements are not compliant. Equipment manufactured after September 5, 1978, shall be subject to the additional requirement that all variable controls for technique factors shall be adjusted to alternate settings and reset to the test setting after each measurement. The percent line-voltage regulation shall be within one (± 1) of the mean value for all measurements.
2. For equipment having automatic exposure controls, compliance shall be determined with a sufficient thickness of attenuating material in the useful beam such that the technique factors can be adjusted to provide individual exposures of a minimum of twelve (12) pulses on field emission equipment rated for pulsed operation or no less than one tenth of one (0.1) second per exposure on all other equipment.
4.6.7 Radiation from Capacitor Energy Storage Equipment
A. Radiation emitted from the X-ray tube shall not exceed:
- 1. An air kerma of twenty-six one hundredths (0.26) µGy (three one hundredths (0.03) mR exposure) in one (1) minute at five centimeters (5 cm) from any accessible surface of the diagnostic source assembly, with the beam-limiting device fully open, the system fully charged, and the exposure switch, timer, or any discharge mechanism not activated. Compliance shall be determined by measurements averaged over an area of one hundred square centimeters (100 cm2), with no linear dimensions greater than twenty centimeters (20 cm); and
2. An air kerma of eighty-eight one hundredths (0.88) mGy (one hundred (100) mR exposure) in one (1) hour at one hundred centimeters (100 cm) from the X-ray source, with beam-limiting device fully open, when the system is discharged through the X-ray tube either manually or automatically by use of a discharge switch or deactivation of the input power. Compliance shall be determined by measurements of the maximum air kerma per discharge multiplied by the total projected number of discharges in one (1) hour (duty cycle). The measurements shall be averaged over an area of one hundred square centimeters (100 cm2) with no linear dimension greater than twenty centimeters (20 cm).
4.6.8 Hand-Holding Restrictions for Portable X-ray Systems
- A. Except during veterinary field operations where it is impractical to do so, a tube stand or other mechanical support shall be used for portable X-ray systems, so that the X-ray tube housing assembly need not be hand-held during exposures.
B. Neither the X-ray tube housing nor the collimating device shall be held during an exposure. Exceptions are allowed for Agency-approved devices specifically designed to be hand-held.
4.6.9 [RESERVED]
4.6.10 Beam-on Indicators
The X-ray control shall provide visual indication whenever X-rays are produced. In addition, a signal audible to the operator shall indicate that the exposure has terminated.
4.6.11 [RESERVED]
4.6.12 Field Limitation and Alignment for Mobile, Portable and Stationary General Purpose X-ray Systems
A. Except when spot-film devices are in service, mobile, portable and stationary general purpose radiographic X-ray systems shall meet the following requirements:
- 1. Variable X-ray Field Limitation. A means for stepless adjustment of the size of the X-ray field shall be provided. Each dimension of the minimum field size at a SID of one hundred centimeters (100 cm) shall be equal to or less than five centimeters (5 cm).
2. Visual Definition
- a. Means for visually defining the perimeter of the X-ray field shall be provided. The total misalignment of the edges of the visually defined field with the respective edges of the X-ray field along either the length or width of the visually defined field shall not exceed two percent (2%) of the distance from the source to the center of the visually defined field when the surface upon which it appears is perpendicular to the axis of the X-ray beam.
- b. When a light localizer is used to define the X-ray field, it shall provide an average illuminance of not less than one hundred sixty (160) lux (fifteen (15) footcandles) at one hundred centimeters (100 cm) or at the maximum SID, whichever is less. The average illuminance shall be based on measurements made in the approximate center of each quadrant of the light field.
c. The edge of the light field at one hundred centimeters (100 cm) or at the maximum SID, whichever is less, shall have a contrast ratio, corrected for ambient lighting, of not less than four (4) in the case of beam-limiting devices designed for use on stationary equipment, and a contrast ratio of not less than three (3) in the case of beam-limiting devices designed for use on mobile and portable equipment. The contrast ratio is defined as I1/I2, where I1 is the illuminance three millimeters (3 mm) from the edge of the light field toward the center of the field; and I2 is the illuminance three millimeters (3 mm) from the edge of the light field away from the center of the field. Compliance shall be determined with a measuring aperture of one millimeters (1 mm).
4.6.13 Field Indication and Alignment on Stationary General Purpose X-ray Equipment
A. Except when spot-film devices are in service, stationary general purpose X-ray systems shall meet the following requirements in addition to those prescribed in § 4.6.12 of this Part:
- 1. Means shall be provided to indicate when the axis of the X-ray beam is perpendicular to the plane of the image receptor, to align the center of the X-ray field with respect to the center of the image receptor to within two percent (2%) of the SID, and to indicate the SID to within two percent (2%);
- 2. The beam-limiting device shall numerically indicate the field size in the plane of the image receptor to which it is adjusted;
- 3. Indication of field size dimensions and SIDs shall be specified in centimeters and/or inches and shall be such that aperture adjustments result in X-ray field dimensions in the plane of the image receptor which correspond to those indicated by the beam-limiting device to within two percent (2%) of the SID when the beam axis is indicated to be perpendicular to the plane of the image receptor; and
4. Compliance measurements will be made at discrete SIDs and image receptor dimensions in common clinical use (such as SIDs of one hundred (100), one hundred fifty (150), and two hundred (200) centimeters and/or thirty-six (36”), forty (40”), forty-eight (48”), seventy-two (72”) inches and nominal image receptor dimensions of thirteen (13), eighteen (18), twenty-four (24), thirty (30), thirty-five (35), forty (40), and forty-three (43) centimeters and/or five (5”), seven (7”), eight (8”), nine (9”), ten (10”), eleven (11”), twelve (12”), fourteen (14”), and seventeen (17”) inches) or at any other specific dimensions at which the beam-limiting device or its associated diagnostic X-ray system is uniquely designed to operate.
4.6.14 Linearity
A. The following requirements apply for any fixed X-ray tube potential within the range of forty percent (40%) to one hundred percent (100%) of the maximum rated:
- 1. Equipment Having Independent Selection of X-Ray Tube Current (mA). The average ratios of air kerma to the indicated milliampere-seconds product (mGy/mAs) obtained at any two (2) consecutive tube current settings shall not differ by more than one tenth (0.10) times their sum: X1-X2 0.10 (X1+X2), where X1 and X2 are the average mGy/mAs values obtained at each of two (2) consecutive tube current settings, or at two (2) settings differing by no more than a factor of two (2) where the mA selector provides continuous selection.
- 2. Equipment Having Selection of X-Ray Tube Current-Exposure Time Product (mAs). For equipment manufactured after May 3, 1994, the average ratios of air kerma to the indicated milliampere-seconds product (mGy/mAs) obtained at any two (2) consecutive mAs selector settings shall not differ by more than one tenth (0.10) times their sum: X1-X2 0.10 (X1 +X2), where X1 and X2 are the average mGy/mAs values obtained at any two (2) consecutive mAs selector settings, or at two (2) settings differing by no more than a factor of two (2) where the mAs selector provides continuous selection.
3. Measuring Compliance. Determination of compliance will be based on ten (10) exposures, made within one (1) hour. These two (2) settings may include any two (2) focal spot sizes except where one (1) is equal to or less than forty-five one hundredths of one millimeter (0.45 mm) and the other is greater than forty-five one hundredths of one millimeter (0.45 mm). For purposes of this requirement, focal spot size is the focal spot size specified by the X-ray tube manufacturer. All values for percent line-voltage regulation at any one (1) combination of technique factors shall be within one (± 1) of the mean value for all measurements at these technique factors.
4.6.15 Field Limitation on Radiographic X-ray Equipment Other Than General Purpose Radiographic Systems
- A. X-ray Systems Designed for One Image Receptor Size. Radiographic equipment designed for only one (1) image receptor size at a fixed SID shall be provided with means to limit the field at the plane of the image receptor to dimensions no greater than those of the image receptor, and to align the center of the X-ray field with the center of image receptor to within two percent (2%) of the SID, or shall be provided with means to both size and align the X-ray field such that the X-ray field at the plane of the image receptor does not extend beyond the edge of the image receptor.
B. Other X-ray Systems. Radiographic systems not specifically covered in §§ 4.6.12, 4.6.13, or 4.6.15(A), of this Part which are also designed for use with extraoral image receptors and when used with an extraoral image receptor, shall be provided with means to limit the X-ray field in the plane of the image receptor so that such field does not exceed each dimension of the image receptor by more than two percent (2%) of the SID, when the axis of the X-ray beam is perpendicular to the plane of the image receptor. In addition, means shall be provided to align the center of the X-ray field with the center of the image receptor to within two percent (2%) of the SID, or means shall be provided to both size and alignment the X-ray field such that the X-ray field at the plane of the image receptor does not extend beyond any edge of the image receptor. These requirements may be met with:
- 1. A system which performs in accordance with §§ 4.6.12 and 4.6.13 of this Part; or when alignment means are also provided, may be met with either;
- 2. An assortment of removable, fixed-aperture, beam-limiting devices sufficient to meet the requirement for each combination of image receptor size and SID for which the unit is designed. Each such device shall have clear and permanent markings to indicate the image receptor size and SID for which it is designed; or
3. A beam-limiting device having multiple fixed apertures sufficient to meet the requirement for each combination of image receptor size and SID for which the unit is designed. Permanent, clearly legible markings shall indicate the image receptor size and SID for which each aperture is designed and shall indicate which aperture is in position for use.
4.6.16 Field Limitation and Alignment for Spot-Film Devices
A. The following requirements shall apply to spot-film devices, except when the spot-film device is provided for use with a radiation therapy simulation system:
- 1. Means shall be provided between the source and the patient for adjustment of the X-ray field size in the plane of the image receptor to the size of that portion of the image receptor which has been selected on the spot-film selector. Such adjustment shall be accomplished automatically when the X-ray field size in the plane of the image receptor is greater than the selected portion of the image receptor. If the X-ray field size is less than the size of the selected portion of the image receptor, the field size shall not open automatically to the size of the selected portion of the image receptor unless the operator has selected that mode of operation.
- 2. Neither the length nor width of the X-ray field in the plane of the image receptor shall differ from the corresponding dimensions of the selected portion of the image receptor by more than three percent (3%) of the SID when adjusted for full coverage of the selected portion of the image receptor. The sum, without regard to sign, of the length and width differences shall not exceed four percent (4%) of the SID. On spot film devices manufactured after February 25, 1978, if the angle between the plane of the image receptor and beam axis is variable, means shall be provided to indicate when the axis of the X-ray beam is perpendicular to the plane of the image receptor, and compliance shall be determined with the beam axis indicated to be perpendicular to the plane of the image receptor.
- 3. The center of the X-ray field in the plane of the image receptor shall be aligned with the center of the selected portion of the image receptor to within two percent (2%) of the SID.
4. Means shall be provided to reduce the X-ray field size in the plane of the image receptor to a size smaller than the selected portion of the image receptor such that:
- a. For spot-film devices used on fixed-SID fluoroscopic systems which are not required to, and do not provide stepless adjustment of the X-ray field, the minimum field size, at the greatest SID, does not exceed one hundred twenty-five square centimeters (125 cm2); or
- b. For spot-film devices used on fluoroscopic systems that have a variable SID and/or stepless adjustment of the field size, the minimum field size, at the greatest SID, shall be containable in a square of five centimeters (5 cm) by five centimeters (5 cm).
- 5. A capability may be provided for overriding the automatic X-ray field size adjustment in case of system failure. If it is so provided, a signal visible at the operator’s position shall indicate whenever the automatic X-ray field size adjustment override is engaged. Each such system failure override switch shall be clearly labeled as follows: FOR X-RAY FIELD LIMITATION SYSTEM FAILURE.
A. Applicability. X-ray systems subject to § 4.7 of this Part shall also meet the applicable requirements of § 4.4 of this Part. Unless otherwise specified, the requirements for equipment contained in § 4.7.1 of this Part are applicable to CT X-ray systems manufactured or remanufactured on or after September 3, 1985.
- 1. Accreditation. All diagnostic CT X-ray equipment for human use shall be accredited by an accrediting organization recognized by the Agency unless otherwise authorized by the Agency.
- 2. Technical and Safety Information. The technical and safety information relating to the conditions of operation, dose information and imaging performance provided by the CT manufacturer shall be maintained by the facility.
B. Termination of Exposure
- 1. Means shall be provided to terminate the X-ray exposure automatically by either de-energizing the X-ray source or shuttering the X-ray beam in the event of equipment failure affecting data collection. Such termination shall occur within an interval that limits the total scan time to no more than one hundred ten percent (110%) of its preset value through the use of either a backup timer or devices which monitor equipment function.
- 2. A visible signal shall indicate when the X-ray exposure has been terminated through the means required by § 4.7.1(B) of this Part.
- 3. The operator shall be able to terminate the X-ray exposure at any time during a scan, or series of scans under CT system control, of greater than one half (0.5) second duration.
C. Tomographic Plane Indication and Alignment
- 1. For any single tomogram system, means shall be provided to permit visual determination of the tomographic plane or a reference plane offset from the tomographic plane.
- 2. For any multiple tomogram system, means shall be provided to permit visual determination of the location of a reference plane. This reference plane can be offset from the location of the tomographic planes.
- 3. If a mechanism using a light source is used to satisfy §§ 4.7.1(C)(1) or (2) of this Part, the light source shall provide illumination levels sufficient to permit visual determination of the location of the tomographic plane or reference plane under ambient light conditions of up to five hundred (500) lux.
D. Beam-On and Shutter Status Indicators and Control Switches
- 1. The CT X-ray control and gantry shall provide visual indication whenever X-rays are produced and, if applicable, whether the shutter is open or closed.
- 2. Each emergency button or switch shall be clearly labeled as to its function.
- E. Indication of CT Conditions of Operation. The CT X-ray system shall be designed such that the CT conditions of operation to be used during a scan or a scan sequence shall be indicated prior to the initiation of a scan or scan sequence. On equipment having all or some of these conditions of operation at fixed values, this requirement may be met by permanent markings. Indication of CT conditions of operation shall be visible from any position from which scan initiation is possible.
F. Additional Requirements Applicable to CT X-Ray Systems Containing a Gantry
- 1. The total error in the indicated location of the tomographic plane or reference plane shall not exceed five millimeters (5 mm).
- 2. If the X-ray production period is less than one half (0.5) second, the indication of X-ray production shall be actuated for at least one half (0.5) second. Indicators at or near the gantry shall be discernible from any point external to the patient opening where insertion of any part of the human body into the primary beam is possible.
- 3. The deviation of indicated scan increment versus actual increment shall not exceed plus or minus one millimeter (± 1 mm) with any mass from zero (0) to one hundred (100) kilograms resting on the support device. The patient support device shall be incremented from a typical starting position to the maximum incremented distance or thirty centimeters (30 cm), whichever is less, and then returned to the starting position. Measurement of actual versus indicated scan increment may be taken anywhere along this travel.
4. Premature termination of the X-ray exposure by the operator shall necessitate resetting of the CT conditions of operation prior to the initiation of another scan.
4.7.2 CT Facility Design Requirements
- A. Aural Communication. Provision shall be made for two (2) way aural communication between the patient and the operator at the control panel.
B. Viewing Systems
- 1. Windows, mirrors, closed-circuit television, or an equivalent shall be provided to permit continuous observation of the patient during irradiation and shall be so located that the operator can observe the patient from the control panel.
2. When the primary viewing system is by electronic means, an alternate viewing system (which may be electronic) shall be available for use in the event of failure of the primary viewing system.
4.7.3 Radiation Protection Survey, Routine Quality Control, and Operating Procedures
A. Radiation Protection Survey
- 1. All CT X-ray systems shall have a radiation protection survey completed by, or under the direct supervision of, a Qualified Medical Physicist within thirty (30) days of installation. In addition, such surveys shall be done after any change in the facility or equipment which might cause a significant increase in radiation hazard.
- 2. The registrant shall obtain a written report of the survey from the Qualified Medical Physicist, and a copy of the report shall be made available to the Agency upon request.
B. Routine Quality Control. A routine QC program on the CT system shall:
- 1. Be developed by a Qualified Medical Physicist and include acceptable tolerances for points evaluated.
- 2. Incorporate the use of a water equivalent phantom. At a minimum, noise, CT number, and artifacts shall be evaluated.
- 3. Be completed at time intervals and under system conditions specified by the Qualified Medical Physicist. The interval shall not exceed one (1) week.
- 4. The registrant shall maintain a record of each routine QC check required by § 4.7.3(B) of this Part for three (3) years.
C. Operating Procedures
- 1. The CT X-ray system shall not be operated except by an individual who has been specifically trained on the operational features of the unit by a manufacturer's applications specialist, Qualified Medical Physicist, or someone deemed qualified by the Agency.
2. The following information shall be readily available to the CT operator:
- a. The results of at least the most recent routine QC completed on the system:
- b. Instructions on performing routine QC, including the use of the CT phantom(s), a schedule of routine QC appropriate for the system, and allowable variations set by the Qualified Medical Physicist for the indicated parameters; and
- c. Scanning protocols established by the RPC, including instructions on reporting deviations.
3. If the Qualified Medical Physicist evaluation or routine QC of the CT X-ray system identifies that a system operating parameter has exceeded a tolerance established by a Qualified Medical Physicist, use of the CT X-ray system on patients shall be limited to those uses permitted by established written instructions of the Qualified Medical Physicist.
4.7.4 CT X-ray System Used for Treatment Planning in Radiation Oncology
- A. A CT X-ray system used solely for treatment planning in radiation oncology is exempt from the specific requirements of §§ 4.7.1, 4.7.2, and 4.7.3 of this Part, and is only subject to the requirements of § 5.10 of this Subchapter.
B. A CT X-ray system used for both diagnostic X-ray and radiation therapy simulation is subject to the requirements of both § 4.7 of this Part and § 5.10 of this Subchapter.
4.7.5 CT Radiation Protocol Committee (RPC)
A. The registrant shall develop and maintain an RPC in accordance with the following:
1. Members of the RPC shall include but not be limited to the:
- a. Lead CT radiologist;
- b. Lead CT technologist;
- c. Qualified Medical Physicist; and
- d. Other individuals as deemed necessary by the registrant (e.g., Radiation Safety Officer, Chief Medical or Administrative Officer, Radiology Department Administrator/Manager).
- 2. If the registrant has more than one (1) site with CT, the registrant may establish a system-wide RPC.
- 3. Two (2) or more registrants may form a cooperative RPC as long as each facility has a representative on the committee.
- 4. If the registrant has already established a radiation safety committee, the requirements of § 4.7.5 of this Part may be delegated to that committee if the members meet the requirements of § 4.7.5(A)(1) of this Part.
B. Responsibilities of the RPC. The RPC shall:
- 1. Review existing CT protocols along with the evaluation and implementation of new and innovative technologies that can improve image quality and/or lower patient dose in comparison with the older protocol.
- 2. Review the capabilities of the individual CT scanner to ensure maximum performance is achieved
3. Determine and review the protocols that are used frequently or could result in significant doses. This review shall include acquisition and reconstruction parameters, image quality, and radiation dose. At a minimum, the facility shall review the following clinical protocols, if performed, at intervals not to exceed twelve (12) months:
- a. Pediatric Head;
- b. Pediatric Abdomen;
- c. Adult Head;
- d. Adult Abdomen;
- e. Adult Chest; and
- f. Brain Perfusion.
4. Establish and implement written protocols, or protocols documented in an electronic reporting system, that include but are not limited to the following:
- a. A method to be used to monitor the CT radiation output.
- b. A standardized protocol naming policy.
- c. A DRL, notification value, and alert value for CT procedures reviewed in § 4.7.5(B)(3) of this Part. Notification and alert values may be applied by using trigger values in conformance with NEMA XR-29 or facility-established values and procedures as defined by the Qualified Medical Physicist.
- d. Actions to be taken for cases when the dose alert value was exceeded which may include patient follow-up.
- e. A process determining who has access and authority to make changes to the protocol management systems, including a method to prevent inadvertent or unauthorized modifications to a CT protocol.
- 5. If CT fluoroscopy is performed, the RPC shall establish and implement operating procedures and training designed to minimize patient and occupational radiation exposure.
- 6. Provide an annual report to the radiation safety committee or radiation safety officer, in the absence of a radiation safety committee,
- 7. At a minimum the RPC members in §§ 4.7.5(A)(1)(a) through (c) of this Part shall meet as often as necessary to conduct business but at intervals not to exceed twelve (12) months.
C. Records
- 1. A record of each RPC meeting shall be maintained. The record shall include the date, names of individuals in attendance, minutes of the meeting, and any action taken.
- 2. The registrant shall maintain a record of RPC policies and procedures.
3. The registrant shall maintain a record of radiation output information so the radiation dose may be estimated in accordance with established protocols (e.g., SSDE). The record shall include:
- a. Patient identification;
- b. Type and date of examination;
- c. Identification of the CT system used; and
d. The dose values the CT system provides (e.g., CTDIvol, DLP, SSDE).
4.7.6 PET CT and SPECT CT System
A. A CT X-ray system used solely to calculate attenuation coefficients in nuclear medicine studies shall meet the applicable requirements in §§ 4.7.1, 4.7.2, 4.7.3, 4.7.5 and 4.10.1 of this Part unless otherwise exempted below:
- 1. § 4.7.1(A)(1) of this Part – Accreditation
- 2. In lieu of § 4.10.1(A)(5) of this Part, a Qualified Medical Physicist shall complete a performance evaluation on the CT system following nationally recognized guidelines or those approved by the Agency at intervals not to exceed twelve (12) months.
- 3. In lieu of § 4.7.1(B) of this Part, routine QC checks shall be completed at intervals not to exceed one (1) week. These checks shall be established and documented by a Qualified Medical Physicist following nationally recognized guidelines or those approved by the Agency.
4. § 4.7.3(C)(2)(c) of this Part – Scanning protocols established by the RPC.
4.7.7 Veterinary CT System
A CT X-ray system, including a cone beam computed tomography system (CBCT), used solely in non-human imaging shall meet the requirements of § 4.7.3(A) of this Part [radiation protection surveys] and is otherwise exempt from the standards of § 4.7 of this Part.
4.7.8 Cone Beam Computed Tomography System (CBCT)
A. A CBCT facility shall meet §§ 4.4, 4.6.2, 4.6.5, 4.7.1(A)(2), and 4.7.1(B), (C), (D), and (E) of this Part, as applicable.
- 1. Exemption. A registrant using fluoroscopy systems capable of CBCT shall meet § 4.7.8 of this Part except §§ 4.7.1(A)(2), and 4.7.1(B), (C), (D), and (E) of this Part.
- B. Beam Alignment. The X-ray field in the plane of the image receptor shall not exceed beyond the edge of the image receptor by more than two percent (2%) of the SID, when the axis of the X-ray beam is perpendicular to the plane of the image receptor. In addition, the center of the X-ray field shall be aligned with the center of the image receptor to within two percent (2%) of the SID.
- C. The registrant or Radiation Protocol Committee (RPC), if established, shall implement and document a policy addressing deviations from established protocols.
- D. A CBCT X-ray system shall only be operated by an individual who has been specifically trained in its operation.
E. The following information shall be readily available to the CBCT operator:
- 1. Instructions on performing routine QC, including the use of the CBCT phantom(s);
- 2. A schedule of routine QC appropriate for the system;
- 3. Allowable variations set by the Qualified Medical Physicist, if required, for the indicated parameters; and
- 4. The results of at least the most recent routine QC completed on the system.
In addition to the applicable requirements of §§ 4.3 and 4.4 of this Part, the requirements of § 4.14 of this Part apply to dental facilities using intraoral, panoramic, and cephalometric X-ray equipment. Dental facilities using cone beam computed tomography (CBCT) technology shall follow applicable provisions of § 4.7.8 of this Part.
4.14.2 Warning Label
- A. On systems manufactured on or before June 10, 2006, the control panel containing the main power switch shall bear the following warning statement or the warning statement in § 4.14.2(B) of this Part, legible and accessible to view: "WARNING: This X-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."
B. On systems manufactured after June 10, 2006, the control panel containing the main power switch shall bear the warning statement, legible and accessible to view: "WARNING: This X-ray unit may be dangerous to patient and operator unless safe exposure factors, operating instructions and maintenance schedules are observed."
4.14.3 Radiation Exposure Control
Means shall be provided to initiate the radiation exposure by a deliberate action on the part of the operator, such as the depression of a switch. Radiation exposure shall not be initiated without such an action.
4.14.4 Exposure Control Location and Operator Protection
A. Except for units designed to be hand-held, the exposure control shall allow the operator to be:
- 1. Behind a protective barrier at least two meters (2 m) (six and one half feet (6.5’)) tall or
2. At least two meters (2 m) (six and one half feet (6.5’)) from the tube housing assembly, outside the path of the useful X-ray beam, while making exposures.
4.14.5 Administrative Controls
- A. Patient and image receptor holding devices shall be used when the techniques permit.
- B. Except for units designed to be hand-held, the tube housing and position indicating device (PID) shall not be hand-held during an exposure.
C. Dental fluoroscopy without image intensification shall not be used.
4.14.6 Beam-on Indicators
The X-ray control shall provide visual indication whenever X-rays are produced. In addition, a signal audible to the operator shall indicate that the exposure has terminated.
4.14.7 Multiple Tubes
Where two (2) or more radiographic tubes are controlled by one (1) exposure switch, the tube which has been selected shall be clearly indicated prior to initiation of the exposure. Only the selected tube can be energized. This indication shall be both on the X-ray control panel and at or near the tube housing assembly which has been selected.
4.14.8 Mechanical Support of Tube Head
The tube housing assembly supports shall be adjusted such that the tube housing assembly will remain stable during an exposure unless tube housing movement is a designed function of the X-ray system.
4.14.9 Battery Charge Indicator
On battery-powered generators, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation.
4.14.10 Locks
All position locking, holding, and centering devices on X-ray system components and systems shall function as intended.
4.14.11 Technique Indicators
- A. For X-ray equipment capable of displaying technique factors, the technique factors to be used during an exposure shall be indicated before the exposure begins. If automatic exposure controls are used, the technique factors which are set prior to the exposure shall be indicated.
B. The requirement of § 4.14.11(A) of this Part may be met by permanent markings on equipment having fixed technique factors.
4.14.12 Exposure Reproducibility
For any specific combination of selected technique factors, the estimated coefficient of variation of the air kerma shall be no greater than five one hundredths (0.05).
4.14.13 Timers
Means shall be provided to terminate the exposure at a preset time interval, a preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor.
4.14.14 Kilovolt Peak
Deviation of technique factors from indicated values shall not exceed the limits provided by the manufacturer. (variation of 21 C.F.R. § 1020.31(a)(4)). At a minimum, the kVp on variable kVp units shall be accurate to within ten percent (10%) and within twenty percent (20%) on fixed kVp units.
4.14.15 X-ray Beam Alignment
- A. The useful X-ray beam shall be limited to the area of clinical interest.
B Intraoral Dental Units
- 1. X-ray systems designed for use with an intraoral image receptor shall be provided with means to limit the source-to-skin distance (SSD) to not less than eighteen centimeters (18 cm).
- 2. The X-ray field at the minimum SSD shall be containable in a circle having a diameter of no more than seven centimeters (7 cm).
C. Extraoral, Panoramic and Cephalometric Units
1. X-ray systems designed for use with extraoral image receptors and when used with an extraoral image receptor, shall be provided with means to limit the X-ray field in the plane of the image receptor so that such field does not exceed each dimension of the image receptor by more than two percent (2%) of the SID, when the axis of the X-ray beam is perpendicular to the plane of the image receptor. In addition, means shall be provided to align the center of the X-ray field with the center of the image receptor to within two percent (2%) of the SID, or means shall be provided to both size and alignment of the X-ray field such that the X-ray field at the plane of the image receptor does not extend beyond any edge of the image receptor. These requirements may be met with:
- (a) An assortment of removable, fixed-aperture, beam-limiting devices sufficient to meet the requirement for each combination of image receptor size and SID for which the unit is designed. Each such device shall have clear and permanent markings to indicate the image receptor size and SID for which it is designed; or
(b) A beam-limiting device having multiple fixed apertures sufficient to meet the requirement for each combination of image receptor size and SID for which the unit is designed. Permanent, clearly legible markings shall indicate the image receptor size and SID for which each aperture is designed and shall indicate which aperture is in position for use.
4.14.16 Beam Quality
- A. The Half Value Layer (HVL) of the useful beam for a given X-ray tube potential shall not be less than the values shown in § 4.14.16(B) of this Part [Table 1]. If it is necessary to determine such half-value layer at an X-ray tube potential which is not listed in § 4.14.16(B) of this Part, linear interpolation or extrapolation may be made. Positive means shall be provided to ensure that at least the minimum filtration needed to achieve beam quality requirements is in the useful beam during each exposure. In the case of a system, which is to be operated with more than one (1) thickness of filtration, this requirement can be met by a filter interlocked with the kilovoltage selector which will prevent X-ray emissions if the minimum required filtration is not in place.
B. Table 1 – X-Ray Tube Voltage (kilovolt peak)
| Minimum HVL (mm in Aluminum) |
| Design Operating Range | Measured Operating Potential | Dental X-ray systems designed for use with intraoral image receptors and manufactured after December 1, 1980 | Dental X-ray systems designed for use with intraoral image receptors and manufactured before or on December 1, 1980 | All X-ray systems, except dental X-ray systems designed for use with intraoral image receptors, subject to § 4.14.16 of this Part and manufactured on or after June 10, 2006 |
| Below 51 | 30 | 1.5 | 0.3 | 0.3 |
| 40 | 1.5 | 0.4 | 0.4 |
| 50 | 1.5 | 0.5 | 0.5 |
| 51 to 70 | 51 | 1.5 | 1.2 | 1.3 |
| 60 | 1.5 | 1.3 | 1.5 |
| 70 | 1.5 | 1.5 | 1.8 |
| Above 70 | 71 | 2.1 | 2.1 | 2.5 |
| 80 | 2.3 | 2.3 | 2.9 |
| 90 | 2.5 | 2.5 | 3.2 |
| 100 | 2.7 | 2.7 | 3.6 |
| 110 | 3.0 | 3.0 | 3.9 |
| 120 | 3.2 | 3.2 | 4.3 |
| 130 | 3.5 | 3.5 | 4.7 |
| 140 | 3.8 | 3.8 | 5.0 |
| 150 | 4.1 | 4.1 | 5.4 |
4.14.17 kVp Limitations
Intraoral dental X-ray machines with a measured kVp of less than fifty-one (51) kVp shall not be used to make diagnostic dental radiographs of humans.
4.4.18 Modification of Certified Diagnostic X-ray Components and Equipment
- A. Diagnostic X-ray components and equipment certified in accordance with 21 C.F.R. Part 1020 shall not be modified such that the component or equipment fails to comply with any applicable provision of 21 C.F.R. Part 1020.
B. The owner of a diagnostic X-ray equipment who uses the equipment in a professional or commercial capacity may modify the equipment provided the modification does not result in the failure of the equipment or component to comply with the applicable requirements of 21 C.F.R. Part 1020. The owner who causes such modification need not submit the reports required by 21 C.F.R. Part 1020, provided the owner records the date and the details of the modification in the equipment records and maintains this information, and provided the modification of the X-ray system does not result in a failure to comply with 21 C.F.R. Part 1020.
4.14.19 Leakage Radiation from the Diagnostic Source Assembly
The leakage radiation from the diagnostic source assembly measured at a distance of one meter (1 m) in any direction from the source shall not exceed eighty-eight one hundredths milligray (0.88 mGy) air kerma (vice one hundred milliroentgen (100 mR) exposure) in one (1) hour when the X-ray tube is operated at its leakage technique factors. If the maximum rated peak tube potential of the tube housing assembly is greater than the maximum rated peak tube potential for the diagnostic source assembly, positive means shall be provided to limit the maximum X-ray tube potential to that of the diagnostic source assembly. Compliance shall be determined by measurements averaged over an area of one hundred square centimeters (100 cm 2) with no linear dimension greater than twenty centimeters (20 cm).
4.14.20 Radiation from Components Other Than the Diagnostic Source Assembly
The radiation emitted by a component other than the diagnostic source assembly shall not exceed an air kerma of eighteen (18) microgray (two (2) milliroentgens exposure) in one (1) hour at five centimeters (5 cm) from any accessible surface of the component when it is operated in an assembled X-ray system under any conditions for which it was designed. Compliance shall be determined by measurements averaged over an area of one hundred square centimeters (100 cm 2) with no linear dimension greater than twenty centimeters (20 cm).
4.14.21 Maintaining Compliance
- Diagnostic X-ray systems and their associated components used on humans and certified pursuant to the Federal X-Ray Equipment Performance Standard (21 C.F.R. Part 1020) shall be maintained in compliance with applicable requirements of that standard.