Ind. Admin. Code tit. 410, r. 5-5-7
Authority: IC 16-41-35-26; IC 16-41-35-29
Affected: IC 16-41-35
Sec. 7. (a) The licensee or registrant shall maintain sufficient calibrated and operable radiation survey instruments to make physical radiation surveys as required by 410 IAC 5-5-7 and 410 IAC 5-4-9 of 410 IAC 5. Instrumentation required by this section shall have such a range such that 2 milliroentgens per hour through 1 roentgen per hour can be measured.
(b) Each radiation survey instrument shall be calibrated:
(c) Records of these calibrations shall be maintained for two years after the calibration date for inspection by the board.
(Indiana Department of Health; Rule HRH-2,PT E,Sec E.104; filed May 26, 1978, 3:30 pm: 1 IR 200; filed Feb 29, 1984, 10:10 am: 7 IR 929; readopted filed Jul 11, 2001, 2:23 p.m.: 24 IR 4234; readopted filed May 22, 2007, 1:44 p.m.: 20070613-IR-410070141RFA; readopted filed Sep 11, 2013, 3:19 p.m.: 20131009-IR-410130346RFA; readopted filed Nov 13, 2019, 3:14 p.m.: 20191211-IR-410190391RFA; readopted filed Nov 14, 2025, 11:18 a.m.: 20251210-IR-410240588RFA)