5 CCR 1005-2
Part 1: General
1.1 Purpose and Scope
This rule establishes minimum standards for certification and approval of entities and processes utilized for alcohol and drug testing. This rule is applicable to: samples taken while driving under the influence, driving while impaired, driving with excessive alcohol content; vehicular assaults and vehicular homicides involving an operator while under the influence of alcohol or one or more drugs or both; the testing of samples of blood or other bodily substances from the bodies of pilots in command, motorboat or sailboat operators in command, or drivers and pedestrians fifteen years of age or older who die within four hours after involvement in a crash involving a motor vehicle, a motorboat, a sailboat or an aircraft; and consumption of alcohol by underage persons and records related thereto.
1.2 Based on evidence gathered through testing and evaluation by the Colorado Department of Public Health and Environment and presented to the Board, the Department and the State Board of Health have determined that the results obtained from the Intoxilyzer 5000EN with software 1358.XX installed are scientifically accurate, precise and reliable, and the collection and preservation of a delayed breath alcohol specimen is not required when this device is properly operated as described in these rules and regulations.
1.3 Evidential Breath Alcohol Testing (EBAT) facilities will operate under Part 5 of these rules and regulations until their Intoxilyzer 5000EN software is upgraded. After an EBAT facility’s Intoxilyzer 5000EN software is upgraded, the EBAT facility will operate under Part 6 of these rules and regulations. All EBAT facilities performing direct evidential breath alcohol tests must comply with Part 6 of these rules and regulations by June 30, 2007.
1.4 Testing of delayed breath specimens operates under Part 5 of these rules and regulations. Testing of blood alcohol, blood drugs and urine drugs operates under Parts 5 and 6 of these rules and regulations.
1.5 Definitions
“Alcohol Percent (%)” – grams of ethanol per 100 milliliters of blood or grams of ethanol per 210 liters of breath.
“Appropriate clinical or public safety facility” – provides for the health and safety of a person whose blood is collected (subject) and meets the following criteria: 1) provide for the washing or cleansing of hands of the blood collection personnel, 2) provide a comfortable chair for the subject with arm supports to assure the elbow remains straight and both arms are accessible to the blood collection personnel, 3) have precautions to assure the subject does not fall out of the chair, 4) provide for cot or other reclining surfaces for subjects who prefer to lie down or who have adverse response to the blood collection procedures, 5) provide for the adverse response to blood collection by providing procedures and equipment for subjects who become faint, nauseous, vomit, bleed excessively, or convulse including the provision of drinking water, and 6) provide for the cleaning and disinfect ion of the blood collection area. “Certification” – the official approval by the Department of an evidential breath alcohol test (EBAT) DEVICE, operator, operator instructor or laboratory to function under these rules and regulations. “Certified Instructor” - an employee of any approved law enforcement agency or the Colorado Department of Public Health and Environment who meets the requirements of Section 2.2 Et. Seq . of these regulations.
“Certified Laboratory” - a laboratory certified by the Department to perform analytical testing of bodily fluids for alcohol or other drugs.
“Certified Operator” - an employee of any approved law enforcement agency or the Colorado Department of Public Health and Environment who meets the requirements of Section 2.1 Et.
Seq. of these regulations.
“Delayed Breath Alcohol Specimens” – the saved ethanol or other analytical components of the EBAT specimen(s).
“Department” – refers to The Colorado Department of Public Health and Environment, Laboratory Services Division.
“Evidential” or “Evidentiary” – refers to a sample which, when tested, gives rise to test results that are sufficiently reliable to be admissible as evidence in a court of law. “Evidential Breath Alcohol Test (EBAT)” – is an evidentiary breath alcohol test performed using a certified evidential breath alcohol test device approved by the Department as described by Section 42-4-1301, C.R.S. (2006).
“Evidential Breath Alcohol Test (EBAT) device” – any instrument certified to perform “Evidential” Breath Alcohol Tests as identified in Section 42-4-1301, C.R.S. (2006). The Intoxilyzer 5000EN is the only evidential breath alcohol testing device certified for use in performing evidential breath alcohol tests.
“Facility” – any location that meets the requirements of these regulations and which is approved by the Department to perform evidential breath alcohol testing. “Proficiency Testing” - The evaluation of unknown specimens supplied by a provider which determines target values for those unknown specimens.
“Representative of a certified laboratory” – any employee of a certified laboratory or a courier employed by or contracted by the certified laboratory to transport specimens for the certified laboratory.
Part 2: Certified Operators and Instructors of Evidential Breath Alcohol Test (EBAT) Devices 2.1 Certification of Operators of EBAT Devices to Determine Alcohol Concentration of Breath Specimens.
2.1.1 To initially be certified as an EBAT operator an individual must:
2.1.1.1 be currently employed by a law enforcement agency or the Department;
2.1.1.2 attend a minimum of eight (8) hours of instruction following the Department’s Operator Training Manual;
2.1.1.3 score 80% or greater on a written exam; and
2.1.1.4 complete a comprehensive practical exam as specified in the Department
Operator Training Manual.
2.1.1.5 upon successful completion of the course requirements, a certificate shall be issued by the Department stating the operator’s name, the course instructor(s) and the initial date of certification.
2.1.2 To maintain certification an operator must:
2.1.2.1 proficiently perform without errors, one EBAT following the procedure specified in Appendix 2A of this regulation in the presence of a certified instructor within a 180 day period.
2.1.2.2 the test performed must be a complete EBAT test.
2.1.2.3 the printout obtained from the certification test shall be signed and dated by the certifying operator and the instructor.
2.1.2.4 the printout must be retained by the law enforcement agency as proof that the certification test was performed in accordance with this regulation.
2.1.3 An operator who fails to certify within the 180 day period must:
2.1.3.1 be decertified by the instructor, and
2.1.3.2 must repeat the 8-hour operator course.
2.1.4 Operators who return after being called to active military service may renew their expired certification by completing the following procedure:
2.1.4.1 document proof of active duty (period of absence must not exceed 2 years.);
2.1.4.2 document proof of last operator certification prior to going on active duty;
2.1.4.3 pass the current operator test with a score of 80% or better;
2.1.4.4 proficiently perform without errors, one Evidential Breath Alcohol Test (EBAT) following the procedure specified in Appendix 2A in the presence of a certified instructor;
2.1.4.5 the documented proof of active duty, documented proof of last operator
certification prior to going on active duty, operator test material and print-out of the certification EBAT must be sent to the Department’s Evidential Breath Alcohol Testing Program.
2.1.4.6 upon successful completion of the requirements in Section 2.1.4 Et. Seq. , a certificate shall be issued by the Department indicating the operator name, the agency certified instructor, the date of certification and “Reinstatement After Military Service.”
2.1.5 A facility must retain records showing each certified operator’s date of original certification and all subsequent dates of certification.
2.2 Certification of Operator Instructors of EBAT Devices to Determine Alcohol Concentration of Breath Specimens
2.2.1 To initially be certified as an EBAT instructor an individual must:
2.2.1.1 be currently employed by a law enforcement agency or the Department;
2.2.1.2 be a currently certified EBAT operator;
2.2.1.3 attend a minimum of sixteen (16) hours of instruction provided by the Department using the Instructor Training Manual;
2.2.1.4 score 80% or greater on a written exam; and
2.2.1.5 complete a comprehensive practical exam as specified in the Department
Instructor Training Manual.
2.2.2 Upon successful completion of the course requirements, a certificate shall be issued by the Department stating the instructor’s name, the Department’s course trainer(s) and the initial date of certification.
2.2.3 A certified instructor is also a certified operator and is authorized to train and certify operators of EBAT devices.
2.2.4 To maintain certification an instructor must:
2.2.4.1 participate in teaching one EBAT operator certification class, or
2.2.4.2 pass a written instructor certification examination within a 365-day period.
2.2.5 An instructor who fails to certify in the 365-day period must be decertified by the Department and must repeat the 16-hour instructor course provided by the Department.
2.2.6 EBAT Instructors who return after being called to active military service may renew their expired certification by completing the following procedure:
2.2.6.1 document proof of active duty (period of absence must not exceed 2 years.);
2.2.6.2 document proof of last instructor certification prior to going on active duty;
2.2.6.3 pass the current instructor test with a score of 80% or better; and
2.2.6.4 proficiently perform without errors, one EBAT test following the procedures specified in Appendix 2A in the presence of a certified instructor.
2.2.6.5 the documented proof of active duty, documented proof of last instructor certification prior to going on active duty, instructor test material and print-out of the certification EBAT must be sent to the Department’s Evidential Breath Alcohol Testing Program.
2.2.7 Upon successful completion of the above requirements, a certificate shall be issued by the Department stating the instructor’s name, the agency certifying the instructor, the Department’s Program Manager or designee, the date of certification and “Reinstatement After Military Service.”
2.2.8 A facility must retain records showing each certified instructor’s date of original certification and dates of all classes taught and written exams taken. Part 3 Blood Testing
3.1 Evidential Specimen Collection
3.1.1 Living Persons
3.1.1.1 Evidential Blood specimen(s) must be:
3.1.1.1.1 collected in the presence of the arresting officer or other responsible person who can authenticate the specimens.
3.1.1.1.2 collected by venipuncture by a physician, nurse, paramedic,
3.1.1.1.3 collected only in an appropriate clinical or public safety facility (e.g., hospital, medical clinic, ambulance, police station, fire station or other approved facility). In no event will the collection of blood specimens interfere with the provision of essential medical care or the ready availability of emergency medical services to the public.
3.1.1.1.4 collected using sterile equipment. The skin at the area of puncture
3.1.2 Deceased Persons
3.1.2.1 Collection of specimens from deceased persons is conducted as per Section 42- 4-1304, C.R.S. (2006), by a person whose training and normal duties include the collection of blood specimens from deceased persons.
3.1.3 Living and Deceased Persons
3.1.3.1 After collection, evidential blood specimens must be:
3.1.3.1.1 dispensed or collected directly into two sterile tubes resulting in a
3.1.3.1.2 inverted several times to properly mix the blood with the sodium
3.1.3.1.3 affixed with an identification label and evidence seal.
3.1.3.1.4 shipped to a certified laboratory. If shipment is delayed for more than 72 hours, the specimens must be placed in secured temporary refrigerated storage at less than 8 degrees Centigrade until shipped but not to exceed 7 days.
3.1.3.2 At the Certified Laboratory:
3.1.3.2.1 one tube of blood must be analyzed for the State's test(s). The State's test(s) must be completed within 15 days of collection.
3.1.3.2.2 the second tube of blood must be refrigerated by the certified
3.1.3.2.3 The second specimen may be released if it is requested and receipted
3.1.3.2.4 The second specimen must be analyzed within 15 days of its receipt by
Part 4: Urine Testing
4.1 Evidential Specimen Collection
4.1.1 Living Persons
4.1.1.1 Urine specimen(s) must be collected in the presence of collection personnel who can authenticate the specimen(s).
4.2 Deceased Persons
4.2.1 Collection of specimens from deceased persons is conducted as per Section 42-4-1304, C.R.S. (2006) by a person whose training and normal duties include the collection of urine samples from deceased persons.
4.3 Living and Deceased Persons
4.3.1 Urine specimen(s) must be:
4.3.1.1 collected in a clean container.
4.3.1.2 affixed with an identification label and evidence seal.
4.3.1.3 shipped to a laboratory certified by the Department. If shipment is delayed for more than 72 hours, the specimens must be placed in secured temporary refrigerated storage at less than 8 degrees Centigrade until shipped but not to exceed 7 days.
4.3.2 At the Certified Laboratory:
4.3.2.1 The State’s test must be completed within 15 days of collection.
4.3.2.2 Any remaining specimen(s) must be retained by the laboratory in frozen storage for a period of not less than 12 months unless requested and receipted for by a representative of another Certified Laboratory.
4.3.2.3 The second specimen must be analyzed by a certified laboratory designated by the defendant or defendant’s legal counsel within 15 days of its receipt by the representative of that Certified Laboratory.
Part 5 Evidential Breath Testing - Collection and Testing Procedures Under Intoxilyzer 5000EN Software Prior to Software Upgrade
5.1 Scope
5.1.1 Part 5 establishes minimum standards for certification and approval of entities and processes utilized for alcohol and drug testing prior to the installation of Intoxilyzer 5000EN software revision 1358.XX.
5.2 Evidential Specimen Collection
5.2.1 Breath – Evidential
5.2.1.1 Evidential breath specimens must be analyzed on EBAT devices approved by the Department. Approval or disapproval of EBAT devices will be based on scientific standards of performance established by the Department. The Intoxilyzer 5000EN is the only EBAT device that may be used for evidential breath alcohol testing.
5.2.1.2 The Department must certify each EBAT device initially and annually thereafter.
5.2.1.3 The Department must issue a certificate for each certified EBAT device after initial certification and after each annual certification. The certificate must reflect the EBAT device approved facility name, the EBAT device serial number and the inclusive dates for the certification period. The certificate for EBAT devices placed in approved mobile facilities must also include the vehicle identification number.
5.2.1.4 An evidential breath alcohol test specimen must only be collected and tested by certified EBAT operators or instructors using a certified EBAT device and following the steps outlined in these regulations.
5.2.1.5 Breath specimens consisting of end-expiratory alveolar air are analyzed to determine their ethyl alcohol concentration.
5.2.2 Breath – Delayed
5.2.2.1 A delayed breath alcohol specimen must be collected with each evidential breath alcohol test pursuant to Appendix 1A.
5.2.2.2 Delayed breath alcohol specimens are considered the personal property of the defendant and retained by the facility for 12 months from the date of collection unless requested and receipted for by a representative of another Certified Laboratory.
5.3 Methods of Analysis
5.3.1 Alcohol in Evidential Breath Specimens
5.3.1.1 The checklist for Evidential Breath Alcohol Tests must be followed as found in Appendix 1A.
5.3.1.2 A system blank(s) analysis must be used with each EBAT.
5.3.1.3 For each EBAT, a Department certified reference standard(s) of known ethanol concentration must be used.
5.3.1.4 A completed EBAT is one in which the checklist contained within Appendix 1A is followed and a printout obtained.
5.4 Laboratory Analysis of Delayed Breath Specimens
5.4.1 Laboratories must be certified by the Department to provide analysis. Certification is based on successful on-site inspection, successful participation in proficiency testing and ongoing compliance.
5.4.2 Laboratories will be certified to perform tests for delayed breath alcohol.
5.4.3 Laboratories must meet standards of performance as established by these regulations. Standards of performance will include personnel qualifications, standard operating procedure manual, analytical process, proficiency testing, quality control, security, chain of custody, specimen retention, space, records, and results reporting.
5.4.4 Laboratory inspections must be performed prior to initial certification and annually thereafter by Department staff as established by these regulations. A laboratory meeting the certification requirements of these regulations will be issued a certificate initially. Recertification shall be required each July 1.
Part 6: Evidential Breath Testing - Collection and Testing Procedures After Installation of Intoxilyzer 5000EN software revision 1358.XX
6.1 Purpose and Scope
6.1.1 Part 6 establishes minimum standards for certification and approval of entities and processes utilized for alcohol and drug testing after installation of Intoxilyzer 5000EN software revision 1358.XX.
6.2 Evidential Specimen Collection
6.2.1 Breath
6.2.1.1 Evidential breath specimens must be analyzed on EBAT devices approved by the Department. Approval or disapproval of EBAT devices will be based on scientific standards of performance established by the Department. The Intoxilyzer 5000EN is the only EBAT device that may be used for Evidential Breath Alcohol Testing.
6.2.1.2 The Department must certify each EBAT device initially and annually thereafter.
6.2.1.3 The Department must issue a certificate for each certified EBAT device after initial certification and after each annual certification. The certificate must reflect the EBAT device approved facility name, the EBAT device serial number and the inclusive dates for the certification period. The certificate for EBAT devices placed in approved mobile EBAT facilities must also include the vehicle identification number.
6.2.1.4 An evidential breath alcohol test specimen must only be collected and tested by certified EBAT operators or instructors using a certified EBAT device and following the steps outlined in these regulations.
6.2.1.5 Breath specimens consisting of end-expiratory alveolar air are analyzed to determine their ethyl alcohol concentration.
6.2.1.6 Unless otherwise provided by law, the subject must be given a choice of which type of evidential chemical test they wish to take to determine the alcohol concentration in their body (evidential breath alcohol test or evidential blood alcohol test) or they may refuse to take either evidential chemical test. Nothing in this regulation is intended to exempt or exonerate an individual from the penalties proscribed in Sections 42 42-4-1301.1 and 42-4-1301.2, C.R.S, or any other relevant law, for the failure to submit to such test.
6.2.1.7 Before the subject is given the choice of the type of evidential chemical test they will take, the certified operator or instructor will include the following information: “You are required to take, complete or cooperate in completing an evidential chemical test to determine the alcoholic content of your blood or breath (C.R.S. 42-4-1301.1(2) (A) (I). The chemical test you choose is the test you will be taking. You cannot choose a different test later. (C.R.S. 42-4-1301.1(2) (A) (II). If you choose a blood test, two (2) tubes of blood will be drawn. One tube belongs to you and you may have it tested at a Health Department Certified Independent Laboratory of your choice. If you choose a breath test, two (2) breath samples will be analyzed by a certified evidential breath alcohol testing device following an approved standard operating procedure. You will not receive a sample to have independently tested by a certified laboratory.
If you refuse to take, complete or cooperate in completing an evidential chemical test to determine the alcoholic content of your blood or breath your driving privilege may be revoked. (C.R.S. 42-2-126(2)(A)(II))”
6.3 Methods of Analysis
6.3.1 Alcohol in Evidential Breath Specimens
6.3.1.1 The EBAT operator or instructor must follow the procedures specified in these regulations for evidential breath alcohol tests.
6.3.1.2 The EBAT operator or instructor must document compliance with these testing procedures by completion of the Department’s checklist form, which is available in Appendix 2A of these regulations or on the Department’s website.
6.3.1.2.1 Information included in Steps 1 through 7 of Appendix 2A must not be
6.3.1.2.2 Steps 1 through 7 must be performed in the order listed.
6.3.1.3 The certified operator or instructor conducting the EBAT test must initial inside the parentheses to the left of each step (1 through 7). Initialing each step indicates that step is properly completed.
6.3.1.4 Step 1. “Turn power switch on or observe the power switch has been activated. If the EBAT device is in the standby mode, press the start test switch.”
6.3.1.4.1 EBAT devices at approved EBAT facilities must always be powered on.
6.3.1.4.2 When the certified operator or instructor first enters the EBAT room
6.3.1.4.3 If the EBAT device is in the standby mode, press the start test switch.
6.3.1.4.4 If the EBAT device is in the ready mode, instrument display scrolling or flashing and simulator display lit, proceed to Step 2.
6.3.1.5 Step 2. “The subject must remove foreign objects from the nose and mouth including dentures. The subject must be closely and continuously observed for 20 minutes prior to testing to assure no belching, regurgitation or intake of any foreign material by nose or mouth has occurred. If such occurs, another 20 minutes of close and continuous observation must elapse under the same conditions.”
6.3.1.5.1 Check the subject for foreign objects in the nose or mouth including
6.3.1.5.2 During the observation period the operator or instructor must be close enough to the subject to detect any belching, regurgitation or intake of foreign material.
6.3.1.5.3 The observation period must be conducted at the EBAT facility by a
6.3.1.5.4 The observation period must not be conducted in the patrol car while
6.3.1.5.5 Start and stop times for the observation period must be recorded from
6.3.1.6 Step 3. “Verify that the external breath tube and simulator vapor tube are both warm.”
6.3.1.6.1 Touch the external breath tube to ensure that it is warm.
6.3.1.6.2 Touch the simulator vapor tube to ensure that it is warm.
6.3.1.6.3 If either tube is cold to the touch, stop the test and call an intoxilyzer instructor for assistance.
6.3.1.7 Step 4. “Observe the simulator temperature is between 33.8 degrees centigrade and 34.2 degrees centigrade.”
6.3.1.7.1 Allow the simulator to equilibrate for a minimum of ten (10) minutes
6.3.1.8 Step 5. “Press the start test switch.”
6.3.1.8.1 Press the green start test switch to initiate the automated test
6.3.1.9 Step 6. “Follow the instructions and sequence of events as they appear on the EBAT device display.”
6.3.1.9.1 A system blank(s) analysis must be used during the test sequence of
6.3.1.9.2 For each EBAT, Department certified reference standard(s) of known
6.3.1.9.3 The results of analyzing more than one reference standard of the same
6.3.1.9.4 If the + or – 10% calibration correlation is not obtained, the
6.3.1.9.5 For each EBAT, the results of the two subject breath alcohol tests must agree with each other within 0.020 grams of alcohol/210 liters of breath.
6.3.1.9.6 During the two minute period between subject breath tests the subject
6.3.1.9.7 During the two minute period between subject breath tests the subject
6.3.1.9.8 A clean mouth piece will be used each time the subject blows into the
6.3.1.10 Step 7. “Retain all printouts generated by the EBAT device with the DUI packet. (ie., error message printouts)
6.3.1.10.1 The officer conducting the EBAT must sign the checklist(s) and
6.3.1.10.2 All printouts generated by the EBAT device must be retained with the
6.3.1.10.3 All records pertaining to the EBAT specimens must be retained by the
6.3.1.10.4 A new checklist, Appendix 2A, must be filled out for each EBAT performed.
6.3.1.10.4.1 A completed evidential breath alcohol Test (EBAT) is one in which
Part 7: Certification of Laboratories
7.1 Laboratory Analysis of Blood, and Urine Specimens
7.1.1 Laboratories must be certified by the Department to provide analysis. Certification is based on successful on-site inspection, successful participation in proficiency testing and ongoing compliance.
7.1.2 Laboratories will be certified to perform tests for one or more of the following categories: blood alcohol, blood drugs, and urine drugs.
7.1.3 Laboratories must meet standards of performance as established by these regulations. Standards of performance will include personnel qualifications, standard operating procedure manual, analytical process, proficiency testing, quality control, security, chain of custody, specimen retention, space, records, and results reporting.
7.1.4 Laboratory inspections must be performed prior to initial certification and annually thereafter by Department staff as established by these regulations. A laboratory meeting their certification requirements of these regulations will be issued a certificate initially. Recertification shall be required each July 1.
7.2 Initial Application
7.2.1 Laboratory directors must submit to the Department a written request for certification of their laboratory.
7.2.2 The Department will acknowledge request and provide a copy of the rules and regulations.
7.2.3 To be certified, laboratories shall meet all requirements in Part 7 of these regulations and pass an on-site inspection.
7.3 Application for Continued Certification
7.3.1 Annually the laboratory director must provide a completed application (Appendix 2B) to the Department to be considered for continued certification.
7.3.2 Laboratories must be recertified every July 1.
7.3.3 In order to be recertified on July 1, laboratories must submit their applications for continued certification no later than June 1, which is 30 days prior to the date for recertification.
7.3.4 To maintain certification, laboratories shall meet all requirements in Part 7 of these regulations and pass an on-site inspection.
7.4 General Requirements
7.4.1 In addition to the laboratory’s application, the laboratory must provide the following information to the Department: written evidence concerning the education, scientific training, and experience of the laboratory director and personnel performing the testing.
7.4.2 Prior to independently analyzing samples, testing personnel must demonstrate acceptable performance on precision, accuracy, specificity, reportable ranges, blanks, and unknown challenge samples (proficiency samples or internally generated quality controls). The laboratory must have a system to evaluate employee competency at regular intervals, not to exceed 12 months.
7.4.3 The laboratory must notify the Department in writing within thirty days of any changes pertaining to Laboratory location, personnel, and analytical methods.
7.4.4 The laboratory director is directly responsible for the accuracy of the tests performed, the accuracy of the reports issued, and adherence to these regulations.
7.4.5 The laboratory must have adequate space, equipment, materials, and controls available to perform the tests reported.
7.4.5.1 Samples which serve as test controls must be of such quality as could be determined “Certifiable” by National Institute of Standards and Technology (“NIST” ) standards, although such samples need not actually be NIST-Certified. Relevant documentation must be available for inspection.
7.4.5.2 If non-traceable standards are used, the laboratory must establish a system to identify each standard, document its preparation, and data to prove agreement within 5% of the expected value when compared to a NIST traceable standard.
7.4.6 The laboratory must document evidence of the utilization of a written method of analysis (Standard Operating Procedure (SOP)) to perform the tests reported. Critical elements that must be addressed in the SOP are in Appendix 2C, Section B5 (a-t).
7.4.7 The laboratory must demonstrate compliance with these regulations through a successful on-site inspection conducted by Department staff prior to certification. Certified laboratories will be inspected on an annual announced basis. Certified laboratories will be inspected on an unannounced basis to evaluate complaints.
7.4.8 The laboratory must maintain all records related to analysis for a minimum of two years. Records to be maintained include instrument maintenance, quality control and quality assurance of all analyses performed, specimen processing, results and reports of analysis, dates of analysis and the identity of the person performing the analysis. Retained records must be open to inspection by Department personnel.
7.5 Proficiency Testing of Blood and Urine Samples
7.5.1 Proficiency Testing (PT) is the evaluation of unknown specimens supplied by a provider that determines target values for those unknown specimens. PT is required for each approved category.
7.5.2 Prior to initial certification the laboratory must have successfully participated in one proficiency test event within the past 12 months.
7.5.3 To maintain continued laboratory certification, a laboratory must participate in the PT program and maintain satisfactory performance.
7.5.4 PT samples shall be tested by the same procedure used for all samples, including, but not limited to, the same number of replicate analyses, the same standards, same testing personnel and equipment, and all other pertinent factors.
7.5.5 Blood Alcohol
7.5.5.1 The Department will make arrangements to provide blood alcohol PT samples to the laboratories through a PT provider.
7.5.5.2 A laboratory must participate in PT testing through 3 events per year, consisting of 5 specimens each. The laboratory will submit results to the PT provider. The PT provider will evaluate the results and forward them to the laboratory as well as the Department.
7.5.5.3 Other volatile forensically significant interferents, such as acetone or toluene, may be included in one or more samples. The inclusions of interferents determine the laboratory’s capability of differentiating the volatile interferent from ethyl alcohol. Identification of these interfering volatiles will be used as a criterion for acceptable performance.
7.5.5.4 Grading Criteria for Blood Alcohol Proficiency Testing
7.5.5.4.1 Proficiency test results must be returned to the PT provider within the time specified by the PT provider. Results received after the due date will not be graded and will be considered an unsatisfactory performance resulting in a score of 0 for the testing event. The laboratory must contact the PT provider if extenuating circumstances prevent timely response to a PT event.
7.5.5.4.2 The laboratory must investigate any score less than 100% and
7.5.5.4.3 The PT provider will score each event as “satisfactory or
7.5.6 Urine-drugs and Blood-drugs
7.5.6.1 For blood and urine drug analyte screening and confirmation certification a laboratory must successfully participate in the appropriate College of American Pathologists (CAP) proficiency test programs.
7.5.6.1.1 For blood-drug certification the required program is the Forensic
7.5.6.1.2 For urine-drug certification the required program is Urine Drug Testing (Confirmatory) UDC.
7.5.6.2 A satisfactory event score is the identification of 80% of the target analytes present with no false positives.
7.5.6.3 The laboratory must request CAP to mail a consultant copy of their survey results to:
Colorado Department of Public Health and EnvironmentLaboratory Services DivisionCertification Program8100 Lowry BoulevardDenver, CO 80230-6828 7.5.6.4 A laboratory will be suspended from a category if two consecutive unsuccessful PT events occur. A laboratory may be reinstated to active status after successful participation in the next test event. Failure to successfully participate in the next test event will result in the denial, suspension or revocation of the certificate and require two successful PT events before the laboratory may reapply for certification. The laboratory may request the PT provider to send, at the expense of the laboratory, one extra set of PT samples when in suspension status.
7.6 On-Site Laboratory Inspection
7.6.1 On-site laboratory inspections must be performed prior to initial certification and annually thereafter by the Department.
7.6.2 The on-site inspection will include a review of the laboratory’s practices to assure compliance with these regulations. The requirements are in checklist format in Appendix 2C.
7.6.3 Laboratories will be contacted to arrange routine inspection dates approximately three weeks prior to a proposed date. A letter confirming the inspection date will be sent to the laboratory.
7.6.4 The on-site inspection’s checklist will be used systematically to evaluate and assess a laboratory’s compliance. Each item listed on the checklist will be answered by the department inspector as, Yes (“Y” ), No (“N” ) or Not Applicable (“NA” ). Each item answered as “N” will be described in a report to include the noncompliant practice, the source of information and the scope and extent of the noncompliant practice.
7.6.5 Following the on-site inspection, a written report will be prepared and reviewed by a peer inspector or supervisor prior to mailing. The report should be mailed to the laboratory within 15 days of inspection.
7.6.6 The laboratory must provide a written response to the report when noncompliant practices are identified. The laboratory must provide a written plan of correction within 15 days of receipt of the written inspection report for each noncompliant item cited as a result of items marked “N” on the inspection checklist. A response will not be required from the laboratory if all items on an inspection checklist are marked either “Y” or “NA” .
7.6.7 The written plan of correction will be reviewed by the inspector and if appropriate will be approved. Any items requiring clarification will be resolved by phone or written correspondence.
7.6.8 Documents must be provided to the Department by the laboratory within 90 days of the inspection for verification and proof of implementation of the corrections described in the written plan of correction. A subsequent on-site inspection will be conducted if the verification documents are not received, if compliance with corrective actions are difficult to verify by documentation, or if practices subject to correction have significant potential for direct impact on the quality of laboratory results.
7.6.9 Identification of noncompliant practices directly resulting in inaccurate laboratory reports, failure to provide a plan of correction or failure to correct adequately any noncompliant practice may result in inspector’s recommendation to deny initial certification or limit, deny, suspend or revoke the laboratory certificate. Such action shall be governed by section 24-4-105, C.R.S.
7.6.10 A certificate will be issued by the Department to the laboratory to show certification has been approved. The certificate will reflect the laboratory name, location, the analytical categories approved and the effective dates of the certification period. The certification period will not exceed twelve months.
7.6.11 The Department will annually publish a list of certified laboratories.
7.7 Standards for approved permanent, temporary and mobile Evidential Breath Alcohol Testing (EBAT) facilities 7.7.1 Evidential Breath Alcohol Test(s) must be conducted only in facilities that have been approved by the Department.
7.7.2 Department standards for all approved EBAT facilities are specified in these regulations.
7.7.3 All approved EBAT facilities must meet standards of performance as established by this section of these regulations.
7.7.4 Inspections of permanent, temporary and mobile facilities must be performed prior to initial approval and once in a calendar year thereafter by Department staff.
7.7.5 Initial inspections of permanent and temporary EBAT facilities must be conducted by Department staff using sections 7.7.12.1 Et. Seq. to 7.7.12.7 Et. Seq. of these regulations.
7.7.6 Annual, complaint and follow up inspections of permanent and temporary EBAT facilities must be conducted by Department staff using sections 7.7.12.2 Et. Seq. to 7.7.12.8 Et.
Seq. of these regulations.
7.7.7 Initial inspections of mobile EBAT facilities must be conducted by Department staff using sections7.7.12.1 Et. Seq. ; 7.7.12.3 Et. Seq. to 7.7.12.7 Et. Seq. ; and 7.7.12.9 Et.
Seq. of these regulations.
7.7.8 Annual complaint and follow up inspections of mobile EBAT facilities must be conducted by Department staff using sections 7.7.12.3 Et. Seq. to 7.7.12.9 Et. Seq. of these regulations.
7.7.9 Mobile EBAT facilities, the EBAT device and its associated equipment must be brought to the Department each time a facility inspection is required.
7.7.10 An EBAT device that is used in a mobile EBAT facility must not be used at a permanent or temporary facility unless approved by the Department.
7.7.11 An EBAT device that is used in a permanent or temporary facility must not be used at a mobile facility unless approved by the Department 7.7.12 Department inspection procedure for permanent, temporary and mobile Evidential Breath Alcohol Test Facilities
7.7.12.1 Initial approval – permanent, temporary and mobile EBAT facilities
7.7.12.1.1 Facilities must submit a written request to the Department for
7.7.12.1.2 After receipt of the written request for approval, the Department shall supply a copy of these regulations to the requesting facility.
7.7.12.1.3 The facility EBAT device instructor or DUI enforcement officer is
7.7.12.1.4 After the facility is constructed and ready for use, written verification of compliance with the requirement of this section must be sent to the Department by the facility. The written verification must include a letter from a certified electrician that the power line to the EBAT device is an isolated line.
7.7.12.1.5 Department staff must perform an initial facility inspection to verify compliance with the requirements of this section. Subsequent facility inspections must be performed once in a calendar year by department staff.
7.7.12.2 Power requirements – permanent and temporary facilities
7.7.12.2.1 AC line voltage of 120VAC + or – 10%, 60 HZ with a grounded 3
7.7.12.2.2 The power line to the EBAT device must be an isolated line. Written
7.7.12.2.3 A surge protection device approved by the Department must be
7.7.12.2.4 Only the EBAT device and its associated equipment shall be
7.7.12.3 Environment – permanent, temporary and mobile EBAT facilities
7.7.12.3.1 The temperature of the EBAT device facility must be maintained
7.7.12.3.2 The facility must have adequate lighting so the EBAT operator can
7.7.12.3.3 The area around and under the EBAT device must be free of dust and
7.7.12.3.4 The Evidential Breath Alcohol Testing facility must be kept orderly.
7.7.12.3.5 The EBAT device and breath alcohol simulator must be located on the
7.7.12.3.6 The organizer stand must be placed on a sturdy and adequate work
7.7.12.3.7 The EBAT device shall be in a smoke free environment.
7.7.12.3.8 The facility must have adequate ventilation to prevent vapor build up around the EBAT device be ventilated.
7.7.12.3.9 The facility must not be used to store any cleaning compounds or
7.7.12.3.10 The facility must be secure and not readily accessible to
7.7.12.4 Documents – Permanent, temporary and mobile EBAT facilities
7.7.12.4.1 The following documents must be maintained at the EBAT facility with
7.7.12.5 Supplies – permanent, temporary and mobile EBAT facilities
7.7.12.5.1 The following supplies must be maintained at the EBAT facility with
7.7.12.6 Evidential Breath Alcohol device functions – permanent, temporary and mobile EBAT facilities
7.7.12.6.1 EBAT device time and date must be correct.
7.7.12.6.2 External breath tube must be heating.
7.7.12.6.3 Simulator vapor tube must be heating
7.7.12.6.4 EBAT device test sequence must be correct.
7.7.12.6.5 EBAT device certification date on the printout must be the same as
7.7.12.6.6 EBAT device must be connected to an active analog telephone line at
7.7.12.7 Simulator functions – Permanent, temporary and mobile EBAT facilities
7.7.12.7.1 Active simulator
7.7.12.7.2 Backup simulator(s)
7.7.12.8 Records review – Permanent, temporary and mobile EBAT facilities.
7.7.12.8.1 Review of the Standard Solution Log Sheets must show precise
7.7.12.8.2 Review of the Standard Solution Log Sheet must not indicate an
7.7.12.8.3 The Standard Simulator Solution must be changed every 30 days or
7.7.12.8.4 Diagnostic checks must be performed every 30 days and printouts
7.7.12.8.5 Calibration checks must be performed every 7 days and printouts
7.7.12.8.6 All EBAT sequence numbers must be recorded on the Standard
7.7.12.8.7 All records pertaining to Evidential Breath Alcohol Tests must be
7.7.12.9 Additional requirements for mobile EBAT facilities
7.7.12.9.1 Power
7.7.12.9.2 Environment
PART 8: Violations and Remedies
8.1 Violations
8.1.1 It is a violation of these rules and regulations to perform testing without an appropriate certificate.
8.1.2 Violation of these rules and regulations may result in denial, suspension or revocation of certification as outlined in Part 8 of these rules and regulations.
8.1.3 Generally, a violation will not be cited if:
8.1.3.1 The violation was unavoidable to prevent loss of life, personal injury or severe property damage or there were no feasible alternatives, and provided that proper notification was given to the Department.
8.1.3.2 The violations resulted from matters beyond the control of the facility or laboratory, such as equipment failures that were unavoidable by reasonable quality assurance measures or management controls.
8.2 Right to appeal the denial, suspension or revocation of certification.
8.2.1 Any facility, laboratory, operator or operator instructor whose certification is denied, suspended or revoked under these regulations may seek appeal of that determination pursuant to section 24-4-105, C.R.S. (2006).
8.3 Denial, Suspension or Revocation of Certification:
8.3.1 The Department may deny, suspend or revoke the certification of EBAT device(s) located in a facility, the certification of an operator, the certification of an operator instructor or the certification of a laboratory for one or more of the following causes:
8.3.1.1 Falsification of data or other deceptive practices including false statements by omission or commission relevant to the certification process.
8.3.1.2 Gross incompetence or negligent practice.
8.3.1.3 Willful or repeated violation of any lawful rule, regulation or order of the Department or the Board of Health and its officers.
8.3.1.4 Inadequate space, equipment, or methods utilized for testing.
8.3.1.5 Submission of any test results of another person as those of the subject being evaluated.
8.3.1.6 For a laboratory, failure to continuously participate in proficiency testing.
8.3.1.7 For a laboratory, the receipt of two consecutive “unsatisfactory” evaluations, or achievement of an “unsatisfactory” score in 2 of any 3 consecutive proficiency test events.
8.3.1.8 For a laboratory, contact with another laboratory concerning proficiency test results prior to the due date of those results.
8.4 Injunction
8.4.1 The Department may seek an injunction against any entity for failure to comply with these rules and regulations.
APPENDIX 1A TITLE: Checklist for Evidential Breath Alcohol Test(s).
1. The subject must remove foreign objects from the nose and mouth to include dentures. The subject must be closely and continuously observed for 20 minutes prior to testing to assure no belching, regurgitation or intake of any foreign material by nose or mouth has occurred. If such occurs, another 20 minutes of close and continuous observation must elapse under the same conditions.
2. Turn power switch on and/or observe the power switch has been activated.
3. Observe the simulator temperature is between 33.8 degrees centigrade and 34.2 degrees centigrade.
4. Activate the Start Test switch.
5. Follow the instructions and sequence of events as they appear on the device display.
6. After the sequence of events has been completed package and seal the Delayed Breath Alcohol specimen.
7. Record the evidential breath alcohol test information on the standard simulator log sheet. APPENDIX 2A APPENDIX 2B APPENDIX 2C