Wash. Admin. Code § 296-817-40035
(1) You must use pure tone, air conduction, hearing threshold examinations, with test frequencies including as a minimum 500, 1000, 2000, 3000, 4000, and 6000 Hz.
(3) You must check the functional operation of the audiometer each day before use by doing all of the following:
(6) You must provide audiometric test rooms that meet the requirements of ANSI S3.1-1999 American National Standard Maximum Permissible Ambient Noise Levels for Audiometric Test Rooms using the following table of maximum ambient sound pressure levels:
Table 4
Maximum Ambient Sound Pressure Levels
| Frequency (Hz) | 500 | 1000 | 2000 | 4000 | 8000 |
| SoundPressureLevel (dB) | 40 | 40 | 47 | 57 | 62 |
| Note: | The American Industrial Hygiene Association and National Hearing Conservation Association recommend conducting audiograms using the requirements of ANSI S3.1-1999 American National Standard Maximum Permissible Ambient Noise Levels for Audiometric Test Rooms with adjustments at only 500 Hz and below. |
[Statutory Authority: RCW 49.17.010, 49.17.040, 49.17.050. WSR 15-23-086, § 296-817-40035, filed 11/17/15, effective 12/18/15. Statutory Authority: RCW 49.17.010, 49.17.040, 49.17.050, and 49.17.060. WSR 03-11-060, § 296-817-40035, filed 5/19/03, effective 8/1/03.]