(1) A licensee providing mobile remote afterloader service shall:
- (a) Check survey instruments before medical use at each address of use or on each day of use, whichever is more frequent; and
- (b) Account for all sources before departure from a client's address of use.
(2) In addition to the periodic spot-checks required by WAC 246-240-381, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of:
- (a) Electrical interlocks on treatment area access points;
- (b) Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;
- (c) Viewing and intercom systems;
- (d) Applicators, source transfer tubes, and transfer tube-applicator interfaces;
- (e) Radiation monitors used to indicate room exposures;
- (f) Source positioning (accuracy); and
- (g) Radiation monitors used to indicate whether the source has returned to a safe shielded position.
- (3) In addition to the requirements for checks in subsection (2) of this section, a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
- (4) If the results of the checks required in subsection (2) of this section indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
- (5) A licensee shall retain a record of each check required by subsection (2) of this section in accordance with WAC 246-240-626.
[Statutory Authority: RCW 70.98.050. WSR 06-05-019, § 246-240-387, filed 2/6/06, effective 3/9/06.]