A registrant may submit a written request to the department for a variance from WAC 246-230-020(1). The registrant shall not use a security screening system on individuals until the department approves the variance request.
(1) The written request must be addressed to: X-ray Supervisor, Office of Radiation Protection, Department of Health, P.O. Box 47827, Olympia, Washington 98504-7827, and must include:
- (a) An explanation of the circumstances involved, and the reason why a limited-use security screening system must be used;
- (b) A description of how using a limited-use security screening system meets the intent of this chapter and how the registrant will protect individuals and operators;
- (c) A description of the limited-use security screening system to be used with supporting pictures or documents; and
- (d) The time period for which the variance is requested.
- (2) If necessary, the department may require the registrant to submit additional information.
- (3) The department may conduct an on-site variance inspection to verify the information provided, or if it otherwise determines that an inspection is necessary.
- (4) As determined by the department, variances may be permanent or temporary.
- (5) The department may impose conditions that may be necessary to protect human health and safety during the term of the variance in order to approve the request.
- (6) The department may, at any time, revoke a variance approval if it is determined that the terms and conditions of the variance are not being followed.
[Statutory Authority: RCW 70.388.040. WSR 25-01-051, s 246-230-090, filed 12/11/24, effective 1/11/25.]