- (1) The registrant shall confirm that the manufacturer of the security screening system has ensured that operating parameters are optimized for the best performance at the lowest dose.
- (2) The radiation dose delivered to an individual must be ALARA while meeting the required detection performance.
- (3) The total effective dose to an individual must not exceed 0.25 mSV (25 mrem) in a calendar year from security screening systems.
- (4) The area outside the radiation screening zone must not exceed 20 µSv (2 mrem) in any one hour.
[Statutory Authority: RCW 70.388.040. WSR 25-01-051, s 246-230-040, filed 12/11/24, effective 1/11/25.]