Wash. Admin. Code § 246-225A-085
The following requirements apply to hand-held X-ray systems:
(2) The image receptor used with hand-held dental X-ray systems must either be:
(4) Conditions and restrictions using the predicted whole body dose rate (effective dose) to the operator and shallow dose to the fingers:
| If the predicted whole body dose rate (effective dose) is: | And the shallow dose rate to the fingers is: | The operator: | |||
| <100 mrem/yr | <1 rem/yr | Can operate equipment with no additional requirements | |||
| ˃100 mrem/yr | <1 rem/yr | Shall wear a leaded apron of 0.25 mm lead equivalent | |||
| ˃100 mrem/yr | ˃1 rem/yr | • | Shall wear a leaded apron of 0.25 mm lead equivalent; and | ||
| • | Shall use equipment for special needs patients outside of routine dental office settings only |
[Statutory Authority: RCW 70.98.050 and 70.98.080. WSR 11-19-013, § 246-225A-085, filed 9/7/11, effective 10/8/11.]