- (1) In addition to the requirements of Section R313-35-110 veterinary x-ray systems shall meet the requirements of Subsections R313-35-111(2) and R313-35-111(3).
- (2) Warning label. The control panel containing the main power switch shall bear the warning statement, legible and accessible to view: "WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."
- (3) Leakage radiation from the diagnostic source assembly. The leakage radiation from the diagnostic source assembly measured at a distance of one meter in any direction from the source may not exceed 25.8 uC/kg, 100 milliroentgens, in one hour if the x-ray tube is operated at its leakage technique factors.
KEY: industry, x-rays, veterinarians, surveys
Date of Last Change: April 14, 2025
Notice of Continuation: October 19, 2021
Authorizing, and Implemented or Interpreted Law: 19-3-104; 19-6-107