Utah Admin. Code R313-28-56
Deviation of measured technique factors from indicated values of kVp and exposure time shall not exceed the limits specified for that system by its manufacturer. In the absence of manufacturer's specifications, the deviation shall not exceed ten percent of the indicated value for kVp and ten percent of the indicated value for times greater than 50 milliseconds.
KEY: dental, X-rays, mammography, beam limitation
Date of Last Change: September 15, 2025
Notice of Continuation: April 6, 2026
Authorizing, and Implemented or Interpreted Law: 19-3-104; 19-6-107