Utah Admin. Code R313-28-55
Radiation emitted from the x-ray tube when the system is fully charged and the exposure switch or timer is not activated shall not exceed a rate of 0.516 uC/kg (two milliroentgens) per hour at five centimeters from accessible surfaces of the diagnostic source assembly, with the beam-limiting device fully open.
KEY: dental, X-rays, mammography, beam limitation
Date of Last Change: September 15, 2025
Notice of Continuation: April 6, 2026
Authorizing, and Implemented or Interpreted Law: 19-3-104; 19-6-107