Utah Admin. Code R313-28-54
When technique factors, including control panel selections associated with automatic exposure control systems, are held constant the coefficient of variation of exposure for both manual and automatic exposure control systems shall not exceed 0.05. This requirement applies to clinically used techniques.
KEY: dental, X-rays, mammography, beam limitation
Date of Last Change: September 15, 2025
Notice of Continuation: April 6, 2026
Authorizing, and Implemented or Interpreted Law: 19-3-104; 19-6-107