- (1) For direct measurements performed in accordance with Rule 0400-20-07-.30, a licensee shall possess and use instrumentation to measure the activity of unsealed radioactive material before it is administered to each patient or human research subject.
- (2) A licensee shall calibrate the instrumentation required in paragraph (1) of this rule in accordance with nationally recognized standards or the manufacturer’s instructions.
- (3) A licensee shall retain a record of each instrument calibration required by this rule in accordance with Rule 0400-20-07-.87.
Authority: T.C.A. §§ 4-5-201, et seq.; 68-202-101, et seq.; and 68-202-201, et seq. Administrative History: Original rule filed February 22, 2012; effective May 22, 2012.