- (1) A licensee shall retain a record of each spot-check for remote afterloader units required by Rule 0400-20-07-.73 for 3 years.
(2) The record must include, as applicable:
- (a) The date of the spot-check;
- (b) The manufacturer’s name, model number, and serial number for the remote afterloader unit and source;
- (c) An assessment of timer accuracy;
- (d) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit’s computer; and
- (e) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
Authority: T.C.A. §§ 4-5-201, et seq.; 68-202-101, et seq.; and 68-202-201, et seq. Administrative History: Original rule filed February 22, 2012; effective May 22, 2012.