ARSD 44:03:01:11.01
Prior to construction, the facility owner or the facility's shielding plan designer must submit the floor plans, shielding specifications, and equipment arrangement of any new X-radiation producing equipment installation or any modification of existing X-radiation producing equipment installations to the department for review and approval. The plans must show:
(6) The type of examinations or treatments to be performed with the equipment and the anticipated workload of the system in milliampere-minutes per week.
The department may require the applicant to utilize the services of a qualified physicist to determine the shielding requirements prior to the plan review and approval. The approval of the plans by the department does not preclude the requirement for additional modifications should a subsequent change of operating conditions create the possibility of an individual receiving a dose in excess of the limits.
Source: 26 SDR 96, effective January 23, 2000 ; 50 SDR 41, effective October 11, 2023 .
General Authority: SDCL 34-21-4.1 , 34-21-15 .
Law Implemented: SDCL 34-21-4.1 , 34-21-17 .
Prior versions effective: 2000-01-23.