67 Pa. Code § 77.26
(b) Procedures for calibration testing.
(4) The breath test equipment does not pass the calibration procedure and shall be removed from service if one of the following occur:
(i) In the case of a device which can only be read to two digits, one or more of the five simulator tests in any one of the three separate series of tests yields a result which is one of the following:
(ii) In the case of a device which can be read to three digits, if, in any one of the three separate series of five simulator tests one of the following occurs:
(5) An example of a calibration testing procedure for a breath test device capable of being read to the third decimal place is as follows:
| Constant value | Reading | Difference (absolute value) | |
| First | .10 | .103 | .003 |
| Series | .10 | .104 | .004 |
| .10 | .097 | .003 | |
| .10 | .104 | .004 | |
| .10 | .099 | .001 | |
| .015 = .003 | |||
| 5 | |||
| Second | .05 | .049 | .001 |
| Series | .05 | .054 | .004 |
| .05 | .048 | .002 | |
| .05 | .051 | .001 | |
| .05 | .052 | .002 | |
| .010 = .002 | |||
| 5 | |||
| Third | .15 | .149 | .001 |
| Series | .15 | .151 | .001 |
| .15 | .148 | .002 | |
| .15 | .151 | .001 | |
| .15 | .160 | .010 | |
| .015 = .003 | |||
| 5 |
First series—Average deviation = .003 and .003 does not exceed .005 Second series—Average deviation = .002 and .002 does not exceed .005 Third series—Average deviation = .003 and .003 does not exceed .005 but the fifth test of the series yields a result which is more than .009 above the constant value so the device fails the calibration testing.
The provisions of this § 77.26 adopted December 21, 1984, effective December 22, 1984, 14 Pa.B. 4599; readopted January 12, 1990, effective immediately and apply retroactively to December 22, 1989, 20 Pa.B. 211. Immediately preceding text appears at serial pages (140067) to (140069).
Approval of Equipment
When the Intoxilyzer 500 has been approved as a testing device, and when the appellant’s only evidence that the device is not accurate was the bald, unsupported claim that it has a 50% error rate, the appellant failed to establish that the trial court erred in admitting into evidence the appellant’s blood alcohol content. Commonwealth v. Snell, 811 A.2d 581 (Pa. Super. 2002); appeal denied 820 A.2d 162 (Pa. 2003).
A certificate of calibration, when offered into evidence with a certificate of accuracy, shall be presumptive evidence of the accuracy of breath test equipment. A reading acquired from such certified equipment must be considered a reliable test result for purposes of determining blood alcohol content, and for proving guilt beyond a reasonable doubt of a violation of law regulating driving under the influence of alcohol. Commonwealth v. Mongiovi, 521 A.2d 429 (Pa. Super. 1987).
Intoxilyzer test results were admissable even though the intoxilyzer was not calibrated before the test, since this section, requires only that the calibration be performed within 1 year following breath test. Commonwealth v. Williamson, 514 A.2d 917 (Pa. Super. 1986).
The breathalyzer regulations at 67 Pa. Code § § 77.24—77.26 are reasonable, comport with State law and ensure valid test results. Baldinger v. Commonwealth, 509 A.2d 912 (Pa. Cmwlth. 1986).
This section cited in 67 Pa. Code § 77.25 (relating to accuracy inspection tests for Type A equipment).