- (a) A dose monitoring system shall be capable of independently terminating irradiation.
- (b) A primary system shall terminate irradiation when the preselected number of dose monitor units has been detected by the system.
- (c) A secondary dose monitoring system shall terminate irradiation when either 110% of the preselected number of dose monitor units or 10 dose monitor units (whichever is greater) has been detected by the secondary dose monitoring system.
- (d) For new equipment, an indicator on the control panel shall show which dose monitoring system has terminated irradiation.
Source
The provisions of this § 228.69 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.