- (a) Therapy systems shall be provided with radiation detectors in the radiation head.
- (b) New equipment shall be provided with at least two radiation detectors incorporated into two separate dose monitoring systems.
- (c) Existing equipment shall be provided with at least one radiation detector incorporated into a primary dose monitoring system.
(d) The detector in a dose monitoring system shall be:
- (1) Permanently installed and interlocked to prevent incorrect positioning.
- (2) Part of a dose monitoring system that provides readings in dose monitor units which can be used to calculate the absorbed dose at a reference point in the treatment volume.
- (3) Capable of independently monitoring and controlling the useful beam.
(e) For new equipment, the design of dose monitoring systems shall assure that:
- (1) The malfunctioning of one system does not affect the correct functioning of the second system.
- (2) The failure of an element common to both systems which could affect the correct function of both systems terminates irradiation.
(f) A dose monitoring system shall have a legible display at the control panel. For new equipment, a display shall:
- (1) Maintain a reading until intentionally reset to zero.
- (2) Have only one scale and no scale multiplying factors.
- (3) Utilize a design so that increasing dose is displayed by increasing numbers and that the absorbed dose may be accurately determined under all conditions of use.
- (4) Provide that, in the event of a power failure, the dose monitoring information required in this subsection displayed at the control panel at the time of failure shall be retrievable.
Source
The provisions of this § 228.66 adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894.
Cross References
This section cited in 25 Pa. Code § 228.74 (relating to absorbed dose rate).