- (a) When field radiographic operations are performed, a minimum of two radiographic personnel shall be present to operate the X-ray device. At least one of the radiographic personnel shall be qualified as a radiographer. The other individual may be either a radiographer or a radiographer’s assistant.
- (b) Other than a radiographer, or a radiographer’s assistant who is under the personal supervision of a radiographer, an individual may not manipulate the controls or operate the equipment used in industrial radiographic operations.
(c) At each job site, the following shall be supplied by the registrant or licensee:
- (1) The appropriate barrier ropes and warning signs.
- (2) At least one operable, calibrated radiation survey instrument.
- (3) For each worker requiring monitoring, an individual personnel dosimeter that is processed and evaluated by an NVLAP processor.
- (4) An operable, calibrated direct reading dosimeter with a range of zero to 200 milliroentgen (51.6 µC/kg) for each worker requiring monitoring.
- (d) An industrial radiographic operation may not be performed if any of the items in subsection (c) is not available at the job site or is inoperable.