- (a) Therapeutic X-ray systems shall be secured to prevent unauthorized use whenever the system is unattended.
- (b) If a patient is held in position for radiation therapy, mechanical supporting or restraining devices shall be used.
- (c) The tube housing assembly may not be held by an individual during exposures.
- (d) No individual other than the patient may be in the treatment room during irradiation unless protected by a shielded booth.
- (e) Interlocks, on-off beam control mechanisms and safety and warning devices shall be checked and appropriately serviced at least once in a calendar year.