- (a) The calibration of an X-ray system shall be performed at intervals not to exceed 1 year and after a change of replacement of components which could cause a change in the radiation output.
- (b) The calibration of the radiation output of the X-ray system shall be performed by or under the direction of a qualified expert for radiation therapy calibration who is physically present at the facility during the calibration.
- (c) The calibration of the radiation output of an X-ray system shall be performed with a calibrated instrument. The calibration of the instrument shall be traceable to a National standard. The instrument shall have been calibrated within the preceding 2 years.
- (d) Calibrations made under this section shall be made so that the dose at a reference point in soft tissue may be calculated as accurately as possible but with an uncertainty of no greater than 5%.
(e) The calibration of the X-ray system shall include, but is not limited to, the following determinations:
- (1) The exposure rates for each combination of field size, technique factors, filter and treatment distance used.
- (2) The degree of congruence between the radiation field and the field indicated by the localizing device if a device is present.
- (3) An evaluation of the uniformity of the largest radiation field used.
- (f) Records of calibration performed under this section shall be maintained by the registrant for at least 5 years after completion of the calibration.
- (g) A copy of the most recent X-ray system calibration shall be available at the control panel.
Authority
The provisions of this § 221.74 issued under section 302 of the Radiation Protection Act (35 P.S. § 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P.S. § 510-20); amended under sections 301 and 302 of the Radiation Protection Act (35 P.S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P.S. § 510-20).
Source
The provisions of this § 221.74 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial pages (249309) to (249310).
Cross References
This section cited in 25 Pa. Code § 216.1 (relating to purpose and scope); 25 Pa. Code § 216.3 (relating to exemptions); and 25 Pa. Code § 221.75 (relating to spot checks).