- (a) The fluoroscopic imaging assembly shall be provided with a primary protective barrier which intercepts the entire cross section of the useful beam at any source-to-image receptor distance.
- (b) The X-ray tube used for fluoroscopy may not produce X-rays unless a barrier is in position to intercept the useful beam and the imaging device is in place and operable.
- (c) A means shall be provided for stepless (continuous) adjustment of the field size.
- (d) The minimum field size at the greatest source to image receptor distance shall be containable in a square of 5 centimeters by 5 centimeters unless otherwise provided in 21 CFR 1020.32(b) (relating to fluoroscopic equipment).
- (e) Equipment may not be operated at a source to skin distance less than 30 centimeters or as required under 21 CFR 1020.32(g).
- (f) The width of the X-ray field in the plane of the image receptor may not exceed that of the visible area of the image receptor by more than 3% of the source to image receptor distance. The sum of the excess length and the excess width may not be greater than 4% of the source to image receptor distance.
- (g) For rectangular X-ray fields used with a circular image receptor, the error in alignment shall be determined along the length and width dimensions of the X-ray field which passes through the center of the visible area of the image receptor.
- (h) Compliance with subsections (a)—(g) shall be determined with the beam axis perpendicular to the plane of the image receptor.
(i) Spot-film devices shall meet the following additional requirements:
- (1) A means shall be provided between the source and the patient for adjustment of the X-ray field size to the size of the portion of film which has been selected on the spot-film selector.
- (2) The adjustments shall be automatically accomplished except when the X-ray field size in the plane of the film is smaller than that of the film.
- (3) The total misalignment of the edges of the X-ray field with the respective edges of the selected portion of the image receptor along the length or width dimensions of the X-ray field in the plane of the image receptor may not exceed 3% of the source-to image receptor when adjusted for full coverage of the selected portion of the image receptor.
- (4) The sum, without regard to sign, of the misalignment along any two orthogonal dimensions, may not exceed 4% of the source to image receptor distance.
- (5) The center of the X-ray field in the plane of the film shall be aligned with the center of the film within 2% of the source to image receptor distance.
Authority
The provisions of this § 221.36a issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P. S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P. S. § 510-20).
Source
The provisions of this § 221.36a adopted October 2, 1998, effective October 3, 1998, 28 Pa.B. 4894; amended November 16, 2001, effective November 17, 2001, 31 Pa.B. 6282; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3823. Immediately preceding text appears at serial pages (285678) to (285679).
Cross References
This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems); and 25 Pa. Code § 221.43a (relating to mobile fluoroscopes).