25 Pa. Code § 221.25
(a) Diagnostic X-ray systems shall have filtration that satisfies the requirements of Table I. The requirements of this section shall be considered to have been met if it can be demonstrated that the half value layer of the primary beam is not less than that shown in Table II.
| Design Operating Range | Measured Operating Potential | Minimum HVL (mm of Aluminum) | ||
| Specified Dental Systems1 | Other X-Ray Systems2 | Other X-Ray Systems3 | ||
| Below 51 | 30 | 1.5 | 0.3 | 0.3 |
| 40 | 1.5 | 0.4 | 0.4 | |
| 50 | 1.5 | 0.5 | 0.5 | |
| 51 to 70 | 51 | 1.5 | 1.2 | 1.3 |
| 60 | 1.5 | 1.3 | 1.5 | |
| 70 | 1.5 | 1.5 | 1.8 | |
| Above 70 | 71 | 2.1 | 2.1 | 2.5 |
| 80 | 2.3 | 2.3 | 2.9 | |
| 90 | 2.5 | 2.5 | 3.2 | |
| 100 | 2.7 | 2.7 | 3.6 | |
| 110 | 3.0 | 3.0 | 3.9 | |
| 120 | 3.2 | 3.2 | 4.3 | |
| 130 | 3.5 | 3.5 | 4.7 | |
| 140 | 3.8 | 3.8 | 5.0 | |
| 150 | 4.1 | 4.1 | 5.4 |
1 Dental X-ray systems designed for use with intraoral image receptors and manufactured after December 1, 1980.
2 Dental X-ray systems designed for use with intraoral image receptors and manufactured before or on December 1, 1980, and all other X-ray systems subject to this section and manufactured before June 10, 2006.
3 All X-ray systems, except dental X-ray systems designed for use with intraoral image receptors, subject to this section and manufactured on or after June 10, 2006.
Note: Half-value layers for kilovoltages not listed in Table II may be determined by interpolation or extrapolation.
The provisions of this § 221.25 issued and amended under sections 301 and 302 of the Radiation Protection Act (35 P.S. § § 7110.301 and 7110.302); and section 1920-A of The Administrative Code of 1929 (71 P.S. § 510-20).
The provisions of this § 221.25 adopted December 18, 1987, effective December 19, 1987, 17 Pa.B. 5235; amended July 16, 2004, effective July 17, 2004, 34 Pa.B. 3238; amended October 26, 2018, effective January 24, 2019, 48 Pa.B. 6791. Immediately preceding text appears at serial pages (304494) to (304496).
This section cited in 25 Pa. Code § 221.35a (relating to fluoroscopic X-ray systems).