(1) A registrant shall perform acceptance testing on the treatment planning system of electronic brachytherapy-related computer systems in accordance with current published recommendations from a recognized national professional association with expertise in electronic brachytherapy (when available). In the absence of an acceptance testing protocol published by a national professional association, the manufacturer’s acceptance testing protocol shall be followed.
(a) Acceptance testing shall be performed by, or under the direct supervision of a Qualified Medical Physicist. At a minimum, the acceptance testing shall include, as applicable, verification of:
- (A) The source-specific input parameters required by the dose calculation algorithm;
- (B) The accuracy of dose, dwell time, and treatment time calculations at representative points;
- (C) The accuracy of isodose plots and graphic displays;
- (D) The accuracy of the software used to determine radiation source positions from radiographic images, if applicable; and
- (E) If the treatment-planning system is different from the treatment-delivery system, the accuracy of electronic transfers of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.
- (2) The position indicators in the applicator shall be compared to the actual position of the source or planned swell positions, as appropriate, at the time of commissioning.
- (3) Prior to each patient treatment regimen, the parameters for the treatment shall be evaluated and approved by a Radiation Therapy Physician and a Qualified Medical Physicist for correctness through means independent of that used for determination of the parameters.