A control panel must be designed to provide:
- (1) An indication of whether electrical power is available at the control panel and if activation of the electronic brachytherapy source is possible;
- (2) An indication of whether X-rays are being produced;
- (3) A means for displaying electronic brachytherapy source accelerating voltage(kV), beam current(µA), pre-set value of radiation dose(cGy) and real time display of dose delivered;
- (4) The means for terminating an exposure at any time; and
- (5) An access locking control device that can prevent unauthorized use of the electronic brachytherapy device.
Statutory/Other Authority
ORS 453.605 - 453.807
Statutes/Other Implemented
ORS 453.605 - 453.807
History
PH 4-2013, f. & cert. ef. 1-29-13