(1) Radiation surveys, as required by OAR 333-120-0200, of all X-ray and Hybrid gauges sufficient to show compliance with OAR 333-122-0101 shall be performed:
- (a) Upon installation of the equipment;
- (b) When reviewing safety devices every six months to insure their proper operation;
- (c) When evaluating annual exposure or dose surveys and personal radiation exposure monitoring devices to ensure that operations are conducted safely;
- (d) Following any change in the initial arrangement, number or type of local components in the system;
- (e) Following any maintenance requiring the disassembly or removal of a local component in the system;
- (f) Any time a visual inspection of the local components in the system reveals an abnormal condition;
- (g) Whenever individual monitoring devices show a significant increase over the previous monitoring period or when the readings are approaching the radiation dose limits specified in OAR 333-120-0100.
- (2) Records of all reviews and surveys shall be maintained in accordance with OAR 333-122-0240 for inspection by the Authority.
Statutory/Other Authority
ORS 453.605 – 453.807
Statutes/Other Implemented
ORS 453.605 – 453.807
History
PH 16-2021, adopt filed 04/21/2021, effective 04/21/2021