(1) All X-ray systems shall have radiation surveys performed, as required by OAR 333-120-0200:
- (a) Upon installation of the equipment;
- (b) Following any change in the initial arrangement, number or type of local components in the system;
- (c) Following any maintenance requiring the disassembly or removal of a local component in the system;
- (d) During the performance of maintenance and alignment procedures if the procedures require the presence of a primary X-ray beam when any local component in the system is disassembled or removed;
- (e) Any time a visual inspection of the local components in the system reveals an abnormal condition; and
- (f) Whenever personnel monitoring devices show a significant increase over the previous monitoring period or the readings are approaching the limits specified in OAR 333-120-0100.
- (2) Radiation levels must not exceed the limits outlined in OAR 333-120-0180.
- (3) The registrant must conduct all surveys with a calibrated and operable radiation survey instrument.
Statutory/Other Authority
ORS 453.605 – 453.807
Statutes/Other Implemented
ORS 453.605 – 453.807
History
PH 16-2021, adopt filed 04/21/2021, effective 04/21/2021