- (1) The licensee must establish and implement a Quality Management program pursuant to OAR 333-116-0125 for PET products, as well as other production and calibration products.
- (2) PET instrumentation and other equipment unique to the PET process must meet all applicable radiation protection standards pursuant to division 120 of these rules.
- (3) Area monitors must be visible and audible to accelerator operators. Monitors must be checked for proper operation daily.
- (4) Wasted targets must be treated as radioactive waste and must be properly dismantled, shielded, stored, and disposed.
- (5) Accelerator shielding design and safety must meet requirements of OAR 333-109-0025.
- (6) Shielding around guide-bends, targets, hot-cells, purification manifolds, etc. must ensure that limits in OAR 333-120-0180 and 333-120-0190 have been met in all areas of beam and nuclide production.
- (7) Security provisions for unauthorized access, janitorial services, maintenance, visitors, tours, and personnel-in-training must conform to requirements in OAR 333-120-0180, 333-120-0250 and 333-120-0260.