(1) A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use must perform full calibration measurements on each unit:
- (a) Before the first medical use of the unit;
(b) Before medical use under the following conditions:
- (A) Whenever spot-check measurements indicate that the output differs by more than five percent from the output obtained at the last full calibration corrected mathematically for radioactive decay;
- (B) Following replacement of the sources or following reinstallation of the gamma stereotactic radiosurgery unit in a new location; and
- (C) Following any repair of the gamma stereotactic radiosurgery unit that includes removal of the sources or major repair of the components associated with the source assembly; and
- (c) At intervals not exceeding one year, with the exception that relative helmet factors need only be determined before the first medical use of a helmet and following any damage to a helmet.
(2) To satisfy the requirement of subsection (1)(a) of this rule, full calibration measurements must include determination of:
- (a) The output within +/-three percent;
- (b) Relative helmet factors;
- (c) Isocenter coincidence;
- (d) Timer accuracy and linearity over the range of use;
- (e) On-off error;
- (f) Trunnion centricity;
- (g) Treatment table retraction mechanism, using backup battery power or hydraulic backups with the unit off;
- (h) Helmet microswitches;
- (i) Emergency timing circuits; and
- (j) Stereotactic frames and localizing devices (trunnions).
- (3) A licensee must use the dosimetry system described in OAR 333-116-0560(1) to measure the output for one set of exposure conditions. The remaining radiation measurements required in subsection (2)(a) of this rule may be made using a dosimetry system that indicates relative dose rates.
- (4) A licensee must make full calibration measurements required by section (1) of this rule must be performed in accordance with published protocols accepted by nationally recognized bodies.
- (5) A licensee must mathematically correct the outputs determined in subsection (2)(a) of this rule at intervals not exceeding one month for cobalt-60 and at intervals consistent with one percent physical decay for all other radionuclides.
- (6) Full calibration measurements required by section (1) of this rule and physical decay corrections required by section (5) of this rule must be performed by the authorized medical physicist.
- (7) A licensee must retain a record of each calibration in accordance with OAR 333-100-0057.
Statutory/Other Authority
ORS 453.635
Statutes/Other Implemented
ORS 453.605 - 453.807
History
PH 52-2018, minor correction filed 02/01/2018, effective 02/01/2018
PH 14-2008, f. & cert. ef. 9-15-08
PH 4-2007, f. & cert. ef. 3-1-07
PH 12-2006, f. & cert. ef. 6-16-06
PH 36-2004, f. & cert. ef. 12-1-04
PH 31-2004(Temp), f. & cert. ef. 10-8-04 thru 4-5-05
PH 3-2003, f. & cert. ef. 3-27-03