(1) The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
- (a) The source-specific input parameters required by the dose calculation algorithm;
- (b) The accuracy of dose, dwell time, and treatment time calculations at representative points;
- (c) The accuracy of isodose plots and graphic displays; and
- (d) The accuracy of the software used to determine sealed source positions from radiographic images.
- (2) Acceptance testing must be performed when new software is installed, for each software revision and when new computer hardware or treatment planning system hardware is installed or repaired.
- (3) Records of acceptance testing must be retained in accordance with OAR 333-100-0057.