- (1) Before releasing a patient or a human research subject from licensee control, a licensee shall survey the patient or the human research subject and the remote afterloader unit with a portable radiation detection survey instrument to confirm that the source(s) has been removed from the patient or human research subject and returned to the safe shielded position.
- (2) A licensee shall retain a record of these surveys in accordance with OAR 333-100-0057.
Statutory/Other Authority
ORS 453.635
Statutes/Other Implemented
ORS 453.605 - 453.807
History
PH 14-2018, minor correction filed 01/30/2018, effective 01/30/2018
PH 14-2008, f. & cert. ef. 9-15-08
PH 4-2007, f. & cert. ef. 3-1-07
PH 12-2006, f. & cert. ef. 6-16-06