- (1) For other than unit dosages, a licensee must possess and use instrumentation to measure the radioactivity of alpha- or beta-emitting radionuclides. A licensee must measure, by direct measurement or by combination of measurements and calculations, the amount of radioactivity in dosages of alpha- or beta-emitting radionuclides prior to administration to each patient or human research subject.
(2) A licensee must develop, implement, and maintain written procedures for use of the instrumentation. At a minimum, a licensee must:
- (a) Perform tests before initial use, and following repair, on each instrument for accuracy, linearity, and geometry dependence, unless it is not appropriate for the use of the instrument; and make adjustments when necessary;
- (b) Perform accuracy annually;
- (c) Perform linearity tests annually over the range of medical use; and
- (d) Check each instrument for constancy and proper operation at the beginning of each day of use.
- (3) Accuracy tests must be performed with source(s) that are traceable to National Institute of Standards and Technology (NIST) or by a supplier who has compared the source to a source that was calibrated by NIST.
- (4) A licensee must retain a record of each check and test required by this rule in accordance with OAR 333-100-0057.