- (1) All mobile or portable radiographic systems shall be provided with means to limit the source-to-skin distance to equal to or greater than 30 cm. This is considered to have been met when the collimator or cone provides the required limits.
- (2) Any device provided to limit the SSD must be durable and securely fastened to the system.
Statutory/Other Authority
ORS 453.605 - 453.807
Statutes/Other Implemented
ORS 453.605 - 453.807
History
PH 12-2006, f. & cert. ef. 6-16-06
HD 1-1991, f. & cert. ef. 1-8-91
HD 4-1985, f. & ef. 3-20-85