- (1) Means shall be provided to present the cumulative on-time of the fluoroscopic tube.
- (2) The maximum cumulative time of the timing device shall not exceed five minutes without resetting.
- (3) A signal audible to the fluoroscopist, or the appropriate operator, shall indicate the completion of any preset cumulative on-time; or if no audible signal is provided, the exposure shall terminate.
Statutory/Other Authority
ORS 453.752 - 453.775
Statutes/Other Implemented
ORS 453.752 - 453.775
History
PH 14-2008, f. & cert. ef. 9-15-08
HD 4-1985, f. & ef. 3-20-85