The source-to-skin distance shall not be less than:
- (1) 38 cm on stationary fluoroscopes manufactured on or after August 1, 1974;
- (2) 35.5 cm on stationary fluoroscopes manufactured prior to August 1, 1974;
- (3) 30 cm on all mobile fluoroscopes; and
- (4) 20 cm for image intensified fluoroscopes used for specific surgical application. The written safety procedures must provide precautionary measures to be adhered to during the use of this device.
Statutory/Other Authority
ORS 453.605 - 453.807
Statutes/Other Implemented
ORS 453.605 - 453.807
History
PH 32-2014, f. 12-22-14, cert. ef. 1-1-15
PH 12-2006, f. & cert. ef. 6-16-06
HD 1-1991, f. & cert. ef. 1-8-91
HD 4-1985, f. & ef. 3-20-85