When required by the Authority, and:
- (1) Prior to construction, the floor plans and equipment arrangement of all new installations, or modifications of existing installations, utilizing X-rays for diagnostic or therapeutic purposes must be submitted to the Authority for review and approval. The required information is as set out in division 120.
- (2) The Authority may require the applicant to utilize the services of a qualified expert to determine the shielding requirements prior to the plan review and approval.
- (3) The approval of such plans shall not preclude the requirement of additional modifications should a subsequent analysis of operating conditions indicate the possibility of an individual receiving a dose in excess of the limits prescribed in division 120.
Statutory/Other Authority
ORS 453.605 - 453.807
Statutes/Other Implemented
ORS 453.605 - 453.807
History
PH 14-2008, f. & cert. ef. 9-15-08
PH 12-2006, f. & cert. ef. 6-16-06
HD 15-1994, f. & cert. ef. 5-6-94
HD 1-1991, f. & cert. ef. 1-8-91