- (a) Authorized personnel. The calibration, or calibration check, may be performed by the operator, a designated member of the law enforcement agency employing the operator, or a third party in accordance with the manufacturer's instructions.
- (b) Calibration check interval and procedure. Alcohol screening devices approved by these rules shall be checked for calibration accuracy annually. Calibration accuracy shall be within (=/-) 0.005 g/210L of the value of the reference method at the time of the calibration check. Calibration checks shall be performed in accordance with the manufacturer's instructions.
- (c) Calibration procedure. Calibrations shall be performed in accordance with the manufacturer's instructions when a calibration check falls outside (+/-) 0.005 g/210L of the value of the reference method at the time of the calibration check.
Added at 34 Ok Reg 847, eff 9-11-17