- (a) Each permittee shall make or cause to be made such surveys as may be necessary to establish compliance with these rules.
- (b) Any survey instruments used to make physical radiation surveys in accordance with 310:281-9-11(a) shall be calibrated and operable.
(c) Each radiation survey instrument shall be calibrated:
- (1) by a person recognized by the Department as competent to perform such service;
- (2) at intervals not to exceed 12 months unless a more restrictive time interval is specified in another subchapter of these rules;
- (3) after each survey instrument repair;
- (4) for the types of radiation used and at energies appropriate for use; and at an accuracy within 20 percent of the true radiation level.
- (d) Records of survey instrument calibrations shall be maintained for inspection by the Department.
Added at 8 Ok Reg 3503, eff 8-15-91 (emergency)
Added at 9 Ok Reg 1643, eff 5-29-92
Amended at 21 Ok Reg 1037, eff 5-13-04
Amended at 38 Ok Reg 1954, eff 9-11-21