(a) General requirements.
- (1) Timers. Means shall be provided to terminate the exposure at a preset time interval, preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. In addition, it shall not be possible to make an exposure when the timer is set to a "zero" or "off" position if either position is provided.
- (2) Reproducibility. With a timer setting of 0.3 second or less, the average exposure period (T) shall be greater than or equal to 5 times the maximum exposure period (Tmax) minus the minimum exposure period (Tmin) when 4 timer tests are performed: T> 5(Tmax - Tmin )
(3) X-Ray control.
- (A) An x-ray control shall be incorporated into each dental x-ray system such that an exposure can be terminated by the operator at any time, except for exposures of 0.5 second or less.
- (B) The exposure switch shall be of the dead-man type.
- (C) Each x-ray control shall be located in such a way as to permit the operator to remain in an area of less than 2 millirems in any 1 hour during the entire exposure.
- (4) Exposure reproducibility. The coefficient of variation shall not exceed 0.10 when all technique factors are held constant. This requirement is met when 4 exposures are made at identical technique factors, the value of the average exposure (E) is greater than or equal to 5 times the maximum exposure (Emax) minus the minimum exposure (Emin): E > 5(Emax - Emin )
(b) Additional requirements for dental intraoral systems.
(1) Source-to-Skin distance (SSD). Dental x-ray systems designed for use with an intraoral image receptor shall be provided with means to limit SSD to not less than:
- (A) 18 centimeters if operable above 50 kVp; or
- (B) 10 centimeters if not operable above 50 kVp.
(2) Field limitation.
- (A) Dental x-ray systems designed for use with an intraoral image receptor shall be provided with means to limit the x-ray beam such that the x-ray field, at the minimum SSD, is containable in a circle having a diameter of no more than 7 centimeters or a rectangle with a longer side of no more than 5cm.
- (B) An open-ended, shielded, beam-indicating device shall be used.
(c) Additional requirements for dental extraoral system field limitation.
- (1) Dental rotational panoramic systems shall be provided with means to limit the x-ray beam to the imaging slit in the transverse axis and shall not exceed a total of 0.5 inch larger than the imaging slit in the vertical axis.
- (2) All other dental extraoral x-ray systems (e.g., cephalometric) shall be provided with means to both size and align the x-ray field so that it does not extend beyond any edge of the image receptor by more than 2 percent of the SID.
Added at 8 Ok Reg 3503, eff 8-15-91 (emergency)
Added at 9 Ok Reg 1643, eff 5-29-92
Amended at 21 Ok Reg 1037, eff 5-13-04
Amended at 38 Ok Reg 1954, eff 9-11-21