Okla. Admin. Code § 252:410-11-41
(a) Communication and viewing requirements. Therapeutic systems with energies of less than 1 MeV that are capable of operating above 50 kVp must provide the operator at the control panel:
(b) Additional requirements for In addition to subsection (a) of this Section, <1 MeV systems capable of operating above 150 kVp must have the following features:(1) all protective barriers must be fixed except for entrance doors or beam interceptors;(2) the control panel must be located outside the treatment room; and(3) the system must have interlocks which require all entrance doors to be closed, including doors to any interior booths, before treatment can be begin or continue. If the radiation beam is interrupted by any door opening, the system must be designed to:(A) reduce the exposure at a distance of 1 meter from the focal spot to less than 100 milliRoentgens per hour; and(B) prevent operation from being resumed until the door is closed and the operator reinitiates irradiation by manual action at the control panel.
(3) the system must have interlocks which require all entrance doors to be closed, including doors to any interior booths, before treatment can be begin or continue. If the radiation beam is interrupted by any door opening, the system must be designed to:
Added at 17 Ok Reg 1136, eff 6-1-00