If approved methodologies do not specify minimum laboratory QA/QC requirements for organic chemistry, the following frequencies shall apply, provided that where possible, surrogates shall be used for all samples:
- (1) A minimum of three concentrations of standards for daily calibration plus one continuing calibration check standard after each ten samples tested inside established control limits; if any analytes are outside the limits, repeat all samples tested after the last acceptable continuing calibration check.
- (2) One reagent blank (carried through preparation) per twenty samples tested (or per batch).
- (3) One matrix spike per twenty samples tested (or per batch).
- (4) One matrix spike duplicate per twenty samples tested (or per batch).