The application shall address the following operational issues:
- (1) A listing of equipment to be used for sample analysis, storage, and reporting.
- (2) A description of the methods, equipment, and instruments used by the applicant laboratory for specific analytes which may be in the form of an SOP manual when required.
(3) A written laboratory QA plan which includes but is not limited to:
- (A) A listing of laboratory personnel, including the laboratory director, which gives the academic training, experience, and analytical and supervisory responsibilities of each; and
- (B) A narrative description of the methods used for sample receipt, storage, and disposal.
- (4) Results of laboratory's two most recent proficiency testing rounds, at least 15 calendar days apart.
- (5) A report of a laboratory evaluation conducted by the DEQ or a DEQ approved assessor within the twelve (12) months prior to the date of filing or, for in-state laboratories only, a letter requesting the DEQ to conduct an on-site evaluation. The evaluation report shall verify data submitted in an application, list any deficiencies and be signed by the DEQ or DEQ approved agent.
- (6) If deficiencies are listed in an evaluation report, the applicant shall submit a corrective action plan which specifies deadlines for implementation and completion of the plan. The DEQ may establish conditions, including compliance schedules, for the applicant's corrective action plan.
- (7) Hours of operation.