N.Y. Comp. Codes R. & Regs. tit. 10, § 16.57
(4) The aluminum equivalent of the total filtration in the useful beam shall not be less than that shown below:
| Operating kVp | Minimum total filtration (Inherent plus added) |
|---|---|
| Below 50 kVp | 0.5 mm aluminum |
| 50-70 kVp | 1.5 mm aluminum |
| Above 70 kVp | 2.5 mm aluminum |
(7) The exposure switch shall be of the dead-man type and shall be provided with a cord sufficiently long that the operator can stand at least six feet from the patient, the X-ray tube and the useful beam.
(b) Use.
If a mobile unit is used routinely in one location it shall be considered a fixed installation, and shall meet the requirements of section 16.56 of this Part.
(c) Conditions for operation of equipment.
(a) Equipment.