N.Y. Comp. Codes R. & Regs. tit. 10, § 16.55
(4) The aluminum equivalent of the total filtration in the useful beam shall not be less than that shown below:
| Operating kVp | Minimum total filter (Inherent plus added) |
|---|---|
| Below 50 kVp | 0.5 mm aluminum |
| 50-70 kVp | 1.5 mm aluminum |
| Above 70 kVp | 2.5 mm aluminum |
(6) Each installation shall be arranged so that the operator can stand at least six feet from the patients, the X-ray tube and the useful beam during exposure. A protective barrier shall be provided when the operator cannot stand at least six feet away from the patient, the X-ray tube and useful beam during exposures.
(b) Conditions for operation of equipment.
(a) Equipment.