N.Y. Comp. Codes R. & Regs. tit. 10, § 16.53
(5) The aluminum equivalent of the total filtration in the useful beam shall not be less than that shown below:
| Operating kVp | Minimum total filter (Inherent plus added) |
|---|---|
| Below 50 kVp | 0.5 mm aluminum |
| 50-70 kVp | 1.5 mm aluminum |
| Above 70 kVp | 2.5 mm aluminum |
(8) The tube head shall remain stationary when placed in the exposure position.
(b) Conditions for operation of equipment.
(6) Gonadal shielding of not less than 0.5 mm lead equivalent shall be used for patients who have not passed the reproductive age during radiographic procedures in which the gonads are in the useful beam.
(c) Special installations.
(1) Panoramic installations are dental installations which consist of a tube head with a collimator providing a narrow (1-2mm) useful beam and an extra-oral film carrier which are interlocked in their motion about the patient.
(i) Equipment.
(c) The aluminum equivalent of the total filtration in the useful beam shall not be less than that shown below:
| Operating kVp | Minimum total filter (Inherent plus added) |
|---|---|
| Below 50 kVp | 0.5 mm aluminum |
| 50-70 kVp | 1.5 mm aluminum |
| Above 70 kVp | 2.5 mm aluminum |
(a) Equipment.