N.M. Code R. § 20.3.6.606
In addition to the provisions of 602 and 603 [Sections 602 and 603 of 20.3.6.602 and 603 NMAC], the requirements of this section apply to x-ray equipment and associated facilities used for dental radiography. Criteria for extraoral dental radiographic systems are covered in 605 [Section 605 of 20.3.6.605 NMAC].
A. Source-to-Skin Distance: X-ray systems designed for use with an intraoral image receptor shall be provided with means to limit source-to-skin distance to not less than:
(1) 18 cm (7.09 inches) if operable above 50 kVp; or
(2) 10 cm (3.93 inches) if not operable above 50 kVp.
B. Field Limitation:
(1) Radiographic systems designed for use with an intraoral image receptor shall be provided with means to limit the x-ray beam such that:
(a) if the minimum source-to-skin distance (SSD) is 18 cm (7.09 inches) or more, the x-ray field, at the minimum SSD, shall be containable in a circle having a diameter of no more than 7 cm (2.76 inches); and
(b) if the minimum SSD is less than 18 cm, (7.09 inches) the x-ray field, at the minimum SSD, shall be containable in a circle having a diameter of no more than 6 cm (2.36 inches).
C. Timers: Means shall be provided to terminate the exposure at a preset time interval, preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. In addition:
(1) Termination of exposure shall cause automatic resetting of the timer to its initial setting or to zero;
(2) It shall not be possible to make an exposure when the timer is to a zero or off position if either position is provided;
(3) Reproducibility: With a timer setting of 0.5 seconds or less, the average exposure period (T) shall be greater than or equal to 5 times the maximum exposure period (Tmax) minus the minimum exposure period (Tmin) when 4 timer tests are performed: T 5(Tmax Tmin).
D. X-ray Control (use of dead-man timers):
(1) A control shall be incorporated into each x-ray system such that an exposure can be terminated at any time, except for exposures of one-half second or less;
(2) Each x-ray control shall be located in such a way as to meet the following criteria:
(a) for stationary x-ray systems, it shall be required that the control switch be permanently mounted in a protected area (e.g., corridor outside the room) so that the operator is required to remain in that protected area during the entire exposure;
(b) for mobile and portable x-ray systems which are:
(i) used for greater than 1 week in one location (one room or suite) shall meet the requirements of 606.D.2.a [Subparagraph (a), Paragraph (2), Subsection D., Section 606 of 20.3.6.606 NMAC];
(ii) used for more than 1 hour and less than 1 week at one location (one room, or suite) shall meet the requirements of 606.D.2.b.(1) [Item (i), Subparagraph (b), Paragraph (2), Subsection D., Section 606 of 20.3.6.606 NMAC]] or be provided with 1.98 m (6.5 feet) high protective barrier which is placed at least 1.83 m (6 feet) from the tube housing assembly and at least 1.83 m (6 feet) from the patient;
(iii) used to make an exposure(s) of only one patient at the use location shall meet the requirement of 606.D.2.b.(1) or 606.D.2.b.(2) [Items (i) or (ii), Subparagraph (b), Paragraph (2), Subsection D., Section 606 of 20.3.6.606 NMAC] or be provided with a method of control which will permit the operator to be a least 3.63 m (12 feet) from the tube head assembly during an exposure.
(3) The x-ray control shall provide visual indication observable at or from the operator's protected position whenever x-rays are produced. In addition, a signal audible to the operator shall indicate that the exposure has terminated; and
(4) From the operator's position, the patient must be capable of being viewed directly or via mirrors.
F. Operating Controls:
(1) Patient and film holding devices shall be used when the techniques permit. The safety rules, required by 602.A.1.d [Subparagraph (d), Paragraph (1), Subsection A., Section 602 of 20.3.6.602 NMAC], shall list individual projections where holding devices cannot be utilized.
(2) Neither the tube housing nor the position indicating device shall be hand-held during an exposure.
(3) The x-ray system shall be arranged and operated in such a manner that the useful beam at the patient's skin does not exceed the dimensions specified in 606.B.1.a or 606.B.1.b [Subparagraphs (a) or (b), Paragraph (1), Subsection B., Section 606 of 20.3.6.606 NMAC].
(4) Dental fluoroscopy shall be prohibited.
[5-3-95; 20.3.6.606 NMAC – Rn, 20 NMAC 3.1.6.606, Recompiled xx/xx/xx]