N.M. Code R. § 20.3.14.1404
C. Each radiation survey instrument shall be calibrated:
(1) by a qualified person or by the manufacturer provided the person or the manufacturer is certified by the Department;
(2) at intervals not to exceed twelve (12) months and after each instrument servicing other than battery replacement; and
(3) to demonstrate an accuracy within plus or minus 20 percent.
D. Records of required calibrations shall be maintained for Department inspection for five years after the calibration date.