N.H. Code Admin. R. He-P 4035.17
Possession, Use, Calibration, and Check of Dose Calibrators
Effective Oct 25, 2022(See Revision Note at part heading for He-P 4035) #6942, eff 2-1-99; ss by #8800, INTERIM, eff 2-1-07, EXPIRED: 7-31-07 New. #8959, eff 8-7-07; ss by #10897, INTERIM, eff 7-22-15, EXPIRES: 1-18-16; ss by #11011, eff 1-15-16; ss by #13048, eff 6-2-20; ss by #13473, eff 10-25-22Former Division of Public Health Services
- (a) For direct measurements performed in accordance with He-P 4035.19, a licensee shall possess and use instrumentation to measure the activity of unsealed byproduct materials before it is administered to each patient or human research subject.
- (b) A licensee shall calibrate the instrumentation required in He-P 4035.17(a) in accordance with nationally recognized standards or the manufacturer’s instructions.
(c) A licensee shall retain a record of each instrument calibration for 3 years. The records shall include:
- (1) The model and serial number of the instrument;
- (2) The date of the calibration;
- (3) The results of the calibration; and
- (4) The name of the individual who performed the calibration.
Source. (See Revision Note at part heading for He-P 4035) #6942, eff 2-1-99; ss by #8800, INTERIM, eff 2-1-07, EXPIRED: 7-31-07 New. #8959, eff 8-7-07; ss by #10897, INTERIM, eff 7-22-15, EXPIRES: 1-18-16; ss by #11011, eff 1-15-16; ss by #13048, eff 6-2-20; ss by #13473, eff 10-25-22