- (a) The laboratory shall meet the instrument calibration requirements specified in Volume 1, Module 2, Section 5.5 of the TNI standards.
- (b) If the requirements in Volume 1, Module 2, Section 5.5 of the TNI standards and the requirements in an approved method differ, the laboratory shall use the requirements that result in a higher level of confidence in testing results.
(c) Prior to analyzing any samples, the laboratory shall verify all initial instrument calibrations with:
- (1) A standard obtained from a manufacturer other than the one that produced the calibration standards used by the laboratory; or
- (2) A standard obtained from the same manufacturer that produced the calibration standards but from a different lot that was independently prepared from different source materials.
- (d) If option (c)(2) is used, the laboratory shall maintain records from the manufacturer to document that the 2 lots were prepared independently from different source materials.
Source. (See Revision Note at chapter heading for Env-C 300) #12065, eff 1-1-17; ss by #13291, eff 11-23-21